Technical data
Chapter 4: Specification & Performance Verification
4-4 2400 Series Operation Manual, 33237 Revision A, May 2004
4.1.3 Spectral Purity
4.1.3.1 Harmonics
6
(Power out = + 6 dBm)
6
Specifications for harmonics above instrument frequency range are typical.
7
Specification is - 30 dBc below 100 MHz.
4.1.3.2 Subharmonics
(Power out = + 6 dBm)
A sub-harmonic is defined as any ¼, ½, or ¾ multiple of the fundamental RF Output
4.1.3.3 Spurious
8
(Offsets > 300 Hz)
8
Specification is - 45 dBc typical for offsets < 300 Hz.
Frequency (GHz) Harmonic (dBc)
0.01 to 2 GHz
7
- 50 dBc
2 to 20 GHz - 55 dBc
20 to 40 GHz - 30 dBc
Frequency (GHz) Harmonic (dBc)
0.01 to 2 GHz - 80 dBc
2 to 20 GHz - 60 dBc
20 to 40 GHz - 50 dBc
Frequency (GHz) Non Harmonics (Offsets > 300 Hz)
0.01 to 16 GHz - 60 dBc
16 to 32 GHz - 54 dBc
32 to 40 GHz - 48 dBc