AX2000 Series Vortex Volumetric and Mass Flow Meters Models AX2200 and AX2300 User’s Manual Azbil Corporation CM2-AX2000-2001 (Revised 04/2014)
AX2200/AX2300 Instruction Manual Table of Contents Customer Notice for Oxygen Service Unless you have specifically ordered Azbil’s optional O2 cleaning, this flow meter may not be fit for oxygen service. Some models can only be properly cleaned during the manufacturing process. Azbil North America, Inc., is not liable for any damage or personal injury, whatsoever, resulting from the use of Azbil North America’s standard mass flow meters for oxygen gas.
Table of Contents AX2200/AX2300 Instruction Manual Table of Contents Chapter 1 Introduction AX2000 Series Vortex Mass Flow Meters ............................................ 1-1 Using this Manual ......................................................................... 1-1 Note and Safety Information ......................................................... 1-2 Receipt of System Components .................................................... 1-2 Technical Assistance ......................................
AX2200/AX2300 Instruction Manual Table of Contents 4-20 mA Output Connections ..................................................... 2-28 Frequency Output Connections ................................................... 2-29 Pulse Output Connections ........................................................... 2-30 Alarm Output Connections.......................................................... 2-32 Remote Electronics Wiring .........................................................
Table of Contents AX2200/AX2300 Instruction Manual Level Two Hidden Diagnostics Values ......................................... 5-4 Analog Output Calibration .................................................................. 5-7 Troubleshooting the Flow Meter ......................................................... 5-8 First Check Items ................................................................................. 5-8 Record Values ......................................................................
AX2200/AX2300 Instruction Manual Table of Contents 2-20. 2-21. 2-22. 2-23. 2-24. 2-25. 2-26. 2-27. 2-28. 2-29. 2-30. 2-31. 2-32. 2-33. 2-34. 2-35. 2-36. 2-37. 2-38. 2-39 2-40 2-41 2-42. 2-43. 2-44. 2-45. 2-46. Non-Isolated Freq. Out. Using External Power Supply........ 2-23 Backlight Using External Power Supply .............................. 2-23 Loop Power Volumetric Flowmeter Junction Box ............... 2-24 Loop Power Mass Flowmeter Junction Box ......................... 2-25 AC Wiring Terminals ....
Table of Contents AX2200/AX2300 Instruction Manual Warnings and Cautions Warning! Consult the flow meter nameplate for specific flow meter approvals before any hazardous location installation. Hot tapping must be performed by a trained professional. U.S. regulations often require a hot tap permit. The manufacturer of the hot tap equipment and/or the contractor performing the hot tap is responsible for providing proof of such a permit.
AX2200/AX2300 Instruction Manual Table of Contents CM2-AX2000-2001 (Rev.
AX2200/AX2300 Instruction Manual Chapter 1 Introduction Chapter 1 Introduction AX2000 Series Multi-Parameter Vortex Mass Flow Meters The azbil AX2200 In-Line and the AX2300 Insertion Vortex Flow Meters provide a reliable solution for process flow measurement. From a single entry point in the pipeline, AX2000 Series meters offer precise measurements of mass or volumetric flow rates.
Chapter 1 Introduction AX2200/AX2300 Instruction Manual Note and Safety Information We use note, caution and warning statements throughout this book to draw your attention to important information. Warning! This statement appears with information that is important to protect people and equipment from damage. Pay very close attention to all warnings that apply to your application. Caution! Note This statement appears with information that is important for protecting your equipment and performance.
AX2200/AX2300 Instruction Manual Chapter 1 Introduction How the AX2000 Series Vortex Mass Flow Meter Operates Figure 1-1. In-Line Vortex Multi-Parameter Mass Flow Meter azbil Series AX2200 and AX2300 Multi-Parameter Vortex Mass Flow Meters use a unique sensor head to monitor mass flow rate by directly measuring three variables–fluid velocity, temperature and pressure. The built-in flow computer calculates the mass flow rate and volumetric flow rate based on these three direct measurements.
Chapter 1 Introduction AX2200/AX2300 Instruction Manual Vortex Shedding Frequency Von Karman vortices form downstream of a shedder bar into two distinct wakes. The vortices of one wake rotate clockwise while those of the other wake rotate counterclockwise. Vortices generate one at a time, alternating from the left side to the right side of the shedder bar. Vortices interact with their surrounding space by over-powering every other nearby swirl on the verge of development.
AX2200/AX2300 Instruction Manual Chapter 1 Introduction Flow Velocity Range To ensure trouble-free operation, vortex flow meters must be correctly sized so that the flow velocity range through the meter lies within the measurable velocity range (with acceptable pressure drop) and the linear range. The measurable range is defined by the minimum and maximum velocity using the following table. Gas Vmin Vmax 0.5 in 0.75 in 1 in 1.
Chapter 1 Introduction AX2200/AX2300 Instruction Manual The Strouhal number is the other dimensionless number that quantifies the vortex phenomenon.
AX2200/AX2300 Instruction Manual Chapter 1 Introduction Minimum Back Pressure A minimum back pressure is required to prevent cavitation in the vortex flow meter when installed in liquid applications. Cavitation is a phenomenon wherein a liquid passing through a low pressure zone no longer can remain in the liquid phase and therefore partially flashes (or “boils”) into its vapor phase. The resulting two-phase flow degrades the liquid calibration accuracy.
Chapter 1 Introduction AX2200/AX2300 Instruction Manual Flow Meter Configurations AX2000 Series Vortex Mass Flow Meters are available in two model configurations: AX2200 In-line flow meter (replaces a section of the pipeline) AX2300 Insertion flow meter (requires a “cold” tap or a “hot” tap into an existing pipeline) Both the in-line and insertion configurations are similar in that they both use identical electronics and have similar sensor heads.
AX2200/AX2300 Instruction Manual Chapter 1 Introduction Line Size / Process Connections / Materials The AX2200 In-line model is built for line sizes ½ through 4 inch wafer or ½ through 8 inch flanged design using ANSI 150, 300, 600, PN16, 40, or 64 class flanges. The AX2300 Insertion model can be used in line sizes 2 inch and greater and is built with a compression fitting or packing gland design using 2 inch NPT, or 2 inch flanged connections (ANSI 150, 300, 600, PN16, 40, or 64 class flanges).
AX2200/AX2300 Instruction Manual Chapter 2 Installation Chapter 2 Installation Installation Overview azbil’s AX2000 Series Vortex Flow Meter installations are simple and straightforward. Both the AX2200 In-Line and AX2300 Insertion type flow meter installations are covered in this chapter. After reviewing the installation requirements given below, see page 2-3 for AX2200 installation instructions. See page 2-6 for AX2300 installation instructions. Wiring instructions begin on page 2-20.
Chapter 2 Installation AX2200/AX2300 Instruction Manual Unobstructed Flow Requirements Select an installation site that will minimize possible distortion in the flow profile. Valves, elbows, control valves and other piping components may cause flow disturbances. Check your specific piping condition against the examples shown below.
AX2200/AX2300 Instruction Manual Chapter 2 Installation AX2200 In-Line Flow Meter Installation Install the AX2200 In-Line Flow Meter between two conventional pipe flanges as shown in Figures 2-3 and 2-4. Table 2-1 provides the recommended minimum stud bolt lengths for wafer-style meter body size and different flange ratings. The meter inside diameter is equal to the same size nominal pipe ID in schedule 80. For example, a 2” meter has an ID of 1.939” (2” schedule 80).
Chapter 2 Installation AX2200/AX2300 Instruction Manual Wafer-Style Flow Meter Installation Install the wafer-style meter between two conventional pipe flanges of the same nominal size as the flow meter. If the process fluid is a liquid, make sure the meter is located where the pipe is always full. This may require locating the meter at a low point in the piping system. Note: Vortex flow meters are not suitable for two-phase flows (i.e., liquid and gas mixtures).
AX2200/AX2300 Instruction Manual Chapter 2 Installation 4. Place the remaining studs between the pipe flanges. Tighten the nuts in the sequence shown in Figure 2-2. Check for leaks after tightening the flange bolts Flange-Style Flow Meter Installation Install the flange-style meter between two conventional pipe flanges of the same nominal size as the flow meter. If the process fluid is a liquid, make sure the meter is located where the pipe is always full.
Chapter 2 Installation AX2200/AX2300 Instruction Manual 2. Seat the meter level and square on the mating connections with the flange stamped with a flow arrow on the upstream side, with the arrow head pointing in the direction of flow. Position a gasket in place for each side. Make sure both gaskets are smooth and even with no gasket material extending into the flow profile. Obstructions in the pipeline will disturb the flow and cause inaccurate measurements. 3. Install bolts in both process connections.
AX2200/AX2300 Instruction Manual Chapter 2 Installation Cold Tap Guidelines Refer to a standard code for all pipe tapping operations. The following tapping instructions are general in nature and intended for guideline purposes only. Caution! When using toxic or corrosive gases, purge the line with inert gas for a minimum of four hours at full gas flow before installing the flow meter. 1. Turn off the flow of process gas, liquid or steam. Verify that the line is not pressurized. 2.
Chapter 2 Installation AX2200/AX2300 Instruction Manual Hot Tap Guidelines Refer to a standard code for all pipe tapping operations. The following tapping instructions are general in nature and intended for guideline purposes only. Warning! Hot tapping must be performed by a trained professional. US. regulations often require a hot tap permit. The manufacturer of the hot tap equipment and/or the contractor performing the hot tap is responsible for providing proof of such a permit. 1.
AX2200/AX2300 Instruction Manual Chapter 2 Installation Figure 2-5. Hot Tap Sequence CM2-AX2000-2001 (Rev.
Chapter 2 Installation AX2200/AX2300 Instruction Manual Flow Meter Insertion The sensor head must be properly positioned in the pipe. For this reason, it is important that insertion length calculations are carefully followed. A sensor probe inserted at the wrong depth in the pipe will result in inaccurate readings. Insertion flow meters are applicable to pipes 2 inch and larger. For pipe sizes ten inches and smaller, the centerline of the meter’s sensing head is located at the pipe’s centerline.
AX2200/AX2300 Instruction Manual Chapter 2 Installation Installing Flow Meters with a Compression Connection* Use the following formula to determine insertion length for flow meters (NPT and flanged) with a compression process connection. The installation procedure is given on the next page. Insertion Length Formula I=S–F–R–t Where: I = Insertion length. S= Stem length – the distance from the center of the sensor head to the base of the enclosure adapter (S = 29.47 inches for standard probes; S = 13.
Chapter 2 Installation AX2200/AX2300 Instruction Manual Insertion Procedure for Meters with a Compression Connection Figure 2-7. Flow Meter with Compression Type Fitting 1. Calculate the required sensor probe insertion length. Caution! The sensor alignment pointer must point downstream, in the direction of flow. 2. Fully retract the stem until the sensor head is touching the bottom of the stem housing. Slightly tighten the compression nut to prevent slippage. 3.
AX2200/AX2300 Instruction Manual Chapter 2 Installation Installing Flow Meters with a Packing Gland Connection* Use the formula below to determine the insertion depth for flow meters (NPT and flanged) equipped with an insertion tool. To install, see the next page for instructions for meters with a permanent insertion tool. For meters with a removable insertion tool, see page 2-15. Insertion Length Formula I = F + R + t – 1.35 Where: I = Insertion length.
Chapter 2 Installation AX2200/AX2300 Instruction Manual Insertion Procedure for Flow Meters with Permanent Insertion Tool Figure 2-9. Flow Meter with Permanent Insertion Tool Caution! The sensor alignment pointer must point downstream, in the direction of flow. Note If line pressure is above 500 psig, it could require up to 25 ft lb of torque to insert the flow meter. Do not confuse this with possible interference in the pipe. 1.
AX2200/AX2300 Instruction Manual Chapter 2 Installation Insertion Procedure for Flow Meters with Removable Insertion Tool Figure 2-10. Flow Meter with Removable Insertion Tool Caution! The sensor alignment pointer must point downstream, in the direction of flow. 1. Calculate the required sensor probe insertion length. Measure from the depth marker arrow down the stanchion and scribe a mark at the calculated insertion depth. 2.
Chapter 2 Installation AX2200/AX2300 Instruction Manual 7. Tighten the packing gland nuts to stop leakage around the stem. Do not torque over 20 ft-lbs. 8. Slide the stem clamp back into position. Torque stem clamp bolts to 15 ft-lbs. Replace the stem clamp nuts and torque to 10-15 ft-lbs. 9. To separate the insertion tool from the flow meter, remove four socket head cap bolts securing the upper and lower retractor brackets. Remove the insertion tool.
AX2200/AX2300 Instruction Manual Chapter 2 Installation Insertion Procedure for Flow Meters with No Insertion Tool (Packing Gland Connection) 1. Calculate the required sensor probe insertion length. Warning! The line pressure must be less than 50 psig for installation. 2. Fully retract the stem until the sensor head is touching the bottom of the stem housing. Remove the two top stem clamp nuts and loosen two stem clamp bolts. Slide the stem clamp away to expose the packing gland nuts.
Chapter 2 Installation AX2200/AX2300 Instruction Manual Adjusting Meter Orientation Depending on installation requirements, you may need to adjust the meter orientation. There are two adjustments available. The first rotates the position of the LCD display/keypad and is available on both in-line and insertion meters. The second is to rotate the enclosure position. This adjustment is only allowed on AX2200 In-Line meters. Display/Keypad Adjustment (All Meters) Figure 2-12.
AX2200/AX2300 Instruction Manual Chapter 2 Installation Enclosure Adjustment (AX2200 Models Only) Figure 2-13. Enclosure Viewing Adjustment To avoid damage to the sensor wires, do not rotate the enclosure beyond 180-degrees from the original position. To adjust the enclosure: 1. Remove power to the flow meter. 2. Loosen the three set screws shown above. Rotate the display to the desired position (maximum 180-degrees). 3. Tighten the three set screws. Restore power to the meter. CM2-AX2000-2001 (Rev.
Chapter 2 Installation AX2200/AX2300 Instruction Manual Loop Power Flow Meter Wiring Connections Warning! To avoid potential electric shock, follow National Electric Code safety practices or your local code when wiring this unit to a power source and to peripheral devices. Failure to do so could result in injury or death. All wiring procedures must be performed with the power off.
AX2200/AX2300 Instruction Manual Chapter 2 Installation 4-20 mA Output Connections The AX2000 Series meter has a single 4-20 mA loop. The 4-20 mA loop current is controlled by the meter electronics. The electronics must be wired in series with the sense resistor or current meter. The current control electronics require 12 volts at the input terminals to operate correctly. The maximum loop resistance (load) for the current loop output is dependent upon the supply voltage and is given in Figure 2-16.
Chapter 2 Installation AX2200/AX2300 Instruction Manual Pulse Output Connections The pulse output is used for a remote counter. When the preset volume or mass (defined in the totalizer settings, see page 3-10) has passed the meter, the output provides a 50 millisecond square pulse. The pulse output requires a separate 5 to 36 VDC power supply. The pulse output optical relay is a normally-open single-pole relay. The relay has a nominal 200 volt/160 ohm rating.
AX2200/AX2300 Instruction Manual Chapter 2 Installation Frequency Output Connections The frequency output is used for a remote counter. It can be scaled to output a 1 to 10 kHz signal proportional to mass or volume flow, temperature, pressure or density. The frequency output requires a separate 5 to 36 VDC power supply; however, there are current and power specifications that must be observed. The output can conduct a current up to 40 mA and can dissipate up to 200 mW.
Chapter 2 Installation AX2200/AX2300 Instruction Manual Remote Electronics Wiring The remote electronics enclosure should be mounted in a convenient, easy to reach location. For hazardous location installations, make sure to observe agency requirements for installation. Allow some slack in the interface cable between the junction box and the remote electronics enclosure. To prevent damage to the wiring connections, do not put stress on the terminations at any time.
AX2200/AX2300 Instruction Manual Chapter 2 Installation Figure 2-23. Loop Power Volumetric Flowmeter Junction Box Sensor Connections Supplied Prior to Jan. 1, 2014 (Wires enter the flow connector from the right side of the connector shown above.) TEMPERATURE SHLD 5&6 RED 6 BLK 6 BLK 5 RED 5 VORTEX BLK 1 RED 1 BLK 2 RED 2 SHLD 1&2 SHLD 3&4 BLK 3 BLK 4 RED 4 RED 3 PRESSURE Figure 2-24. Loop Power Mass Flowmeter Junction Box Sensor Connections CM2-AX2000-2001 (Rev.
Chapter 2 Installation AX2200/AX2300 Instruction Manual High Power Meter Wiring Connections Warning! To avoid potential electric shock, follow National Electric Code safety practices or your local code when wiring this unit to a power source and to peripheral devices. Failure to do so could result in injury or death. All AC power connections must be in accordance with published CE directives. All wiring procedures must be performed with the power off.
AX2200/AX2300 Instruction Manual Chapter 2 Installation Figure 2-26. AC Power Connections Figure 2-27. DC Wiring Terminals DC Power Wiring Caution! The DC wire insulation temperature rating must meet or exceed 85°C (185°F). The DC power wire size must be 20 to 10 AWG with the wire stripped 1/4 inch (7 mm). Connect 18 to 36 VDC (300 mA, 9 W maximum) to the +DC Pwr and –DC Pwr terminals on the terminal block. Torque all connections to 4.43 to 5.31 in-lbs (0.5 to 0.6 Nm). Figure 2-28.
Chapter 2 Installation AX2200/AX2300 Instruction Manual 4-20 mA Output Connections The standard AX2000 Series Flow Meter has a single 4-20 mA loop. Two additional loops are available on the optional communication board. The 4-20 mA loop current is controlled by the meter electronics. The electronics must be wired in series with the sense resistor or current meter. The current control electronics require 12 volts at the input terminals to operate correctly.
AX2200/AX2300 Instruction Manual Chapter 2 Installation Figure 2-31. Non-Isolated 4–20 mA Output Using Meter Input Power Supply Figure 2-32. Isolated 4–20 mA Output Using Meter Provided Power Supply Frequency Output Connections The frequency output is used for a remote counter. It can be scaled to output a 1 to 10 kHz signal proportional to mass or volume flow, temperature, pressure or density.
Chapter 2 Installation AX2200/AX2300 Instruction Manual AC or DC powered meters Figure 2-33. Isolated Frequency Output Using External Power Supply PWR PWR DC Powered meters only Figure 2-34. Non-Isolated Frequency Output Using Input Power Supply + 24 VDC Out - 24 VDC Out AC units only Meter provided DC Power Figure 2-35. Isolated Frequency Output Using Meter Provided Power Supply Pulse Output Connections The pulse output is used for a remote counter.
AX2200/AX2300 Instruction Manual Chapter 2 Installation is needed for the pulse output. Use the second configuration if the voltage at the flow meter power supply is an acceptable driver voltage for the load connected. (Take into account that the current used by the pulse load comes from the meter’s power supply). Use the third configuration if you have an AC powered unit only. In any case, the voltage of the pulse output is the same as the voltage supplied to the circuit. Figure 2-36.
Chapter 2 Installation AX2200/AX2300 Instruction Manual Alarm Output Connections One alarm output (Alarm 1) is included on the standard AX2000 Series Flow Meter. Two or more alarms (Alarm 2 and Alarm 3) are included on the optional communication board. The alarm output optical relays are normally-open single-pole relays. The relays have a nominal 200 volt/160 ohm rating.
AX2200/AX2300 Instruction Manual Chapter 2 Installation Figure 2-41. Isolated Alarm Output Using Meter Provided Power Supply Remote Electronics Wiring The remote electronics enclosure should be mounted in a convenient, easy to reach location. For hazardous location installations, make sure to observe agency requirements for installation. Allow some slack in the interface cable between the junction box and the remote electronics enclosure.
Chapter 2 Installation AX2200/AX2300 Instruction Manual Optional Input Electronics Wiring The meter has two optional input wiring terminals. These can be used to input a Remote or Second RTD input in the case of an Energy Monitoring meter, for the input of a Remote Pressure Transducer, to pass a Contact Closure or for a Remote Density measurement to name a few. In any case, the wiring diagram will be included with the meter if any of the options are specified.
AX2200/AX2300 Instruction Manual Chapter 2 Installation Optional External 4-20 mA Input Wiring The meter is set to have Option 1 used for the external input. Programming menus that pertain to the optional 4-20 mA input are located in the Hidden Diagnostics Menu in Chapter 5. Figure 2-44. External 4-20 mA Input Wiring - External Power Supply Follow the above diagram to wire the external 4-20 mA input into the flow meter using an external power supply.
Chapter 2 Installation AX2200/AX2300 Instruction Manual Figure 2-46. External 4-20 mA Input Wiring - AC Powered Meter Follow the above diagram to wire the external 4-20 mA input into the flow meter using power from the 24 VDC output of an AC powered meter. Optional Contact Closure Input Wiring Figure 2-47. Optional Contact Closure Input Wiring Follow the above diagram to wire an external switch input into the flow meter. The meter is configured to have Option 1 used for the external input.
AX2200/AX2300 Instruction Manual Chapter 3 Operation Chapter 3 Operating Instructions After installing the AX2000 Series Vortex Flow Meter, you are ready to begin operation. The sections in this chapter explain the display/keypad commands, meter start-up and programming. The meter is ready to operate at start up without any special programming. To enter parameters and system settings unique to your operation, see the following pages for instructions on using the setup menus.
Chapter 3 Operation AX2200/AX2300 Instruction Manual Start-Up To begin flow meter operation: 1. Verify the flow meter is installed and wired as described in Chapter 2. 2. Apply power to the meter. At start up, the unit runs a series of selfNote Starting the flow meter or pressing EXIT will always display the Run Mode screens. tests that check the RAM, ROM, EPROM and all flow sensing components. After completing the self-test sequence, the Run Mode screens appear. 3.
AX2200/AX2300 Instruction Manual Chapter 3 Operation CM2-AX2000-2001 (Rev.
Chapter 3 Operation AX2200/AX2300 Instruction Manual Using the Setup Menus 3-4 CM2-AX2000-2001 (Rev.
AX2200/AX2300 Instruction Manual Chapter 3 Operation Programming the Flow Meter 1. Enter the Setup Menu by pressing the ENTER key until prompted for a password. (All outputs are disabled while using the Setup Menus.) 2. Use the keys to select the password characters (1234 is the factory-set password). When the password is correctly displayed, press ENTER to continue. 3. Use the Setup Menus described on the following pages to customize the multiparameter features of your AX2000 Series Flow Meter.
Chapter 3 Operation AX2200/AX2300 Instruction Manual Output Menu Run Mode ENTER Password ENTER Output Menu 4-20 mA Output 1 More > Use keys to access menus < Measure > None Mass Volume *** Energy Temp 1,2 Press Density < 4 mA = xxxx > xxxx < 20mA = xxxx > xxxx < Time Const (Sec) xxxx *** see < Measure > None Mass Volume *** Energy Temp 1,2 Press Density < 4 mA = xxxx > xxxx < 20mA = xxxx > xxxx < Time Const (Sec) xxxx *** see < Measure > None Mass Volume *** Energy Temp 1,2 Press Density < 4
AX2200/AX2300 Instruction Manual Chapter 3 Operation Example for Setting an Output The following shows how to set Output 1 to measure mass flow with 4 mA = 0 lb/hr and 20 mA = 100 lb/hr with a time constant of 5 seconds. (All outputs are disabled while using the Setup Menus.) First, set the desired units of measurement: 1. Use keys to move to the Units Menu (see page 3-12). 2. Press key until Mass Flow Unit appears. Press ENTER. 3. Press key until lb appears in the numerator.
Chapter 3 Operation AX2200/AX2300 Instruction Manual Display Menu ENTER Run Mode Password ENTER Display Menu Use keys to access menus Cycle Time (sec) 0 If Cycle Time is set to zero, manual advance is required Number of Digits 2 Used to set the number of digits displayed after the decimal point Display TC (sec) 1 TC = Display Time constant, used to smooth display MF Vf Te Pr De T Y or N A1 A2 A3 Fl Dt E Y or N MF = Mass Flow Vf = Volume Flow Te = Temperature Pr = Pressure De = Density T = Total
AX2200/AX2300 Instruction Manual Chapter 3 Operation Alarms Menu Example for Setting an Alarm The following shows how to set Relay Alarm 1 to activate if the mass flow rate is greater than 100 lb/hr. You can check the alarm configuration in the Run Mode by pressing the keys until Alarm [1] appears. The lower line displays the mass flow rate at which the alarm activates. Note: all outputs are disabled while using the Setup Menus. First, set the desired units of measurement: 1.
Chapter 3 Operation AX2200/AX2300 Instruction Manual Totalizer #1 Menu Run Mode ENTER Password ENTER Totalizer Menu Use keys to access menus Totaling Inactive Mass Volume Energy Example: Maximum flow rate = 600 gallons per minute (600 gallons per minute = 10 gallons per second) (unit) / Pulse xxxx If unit per pulse is set to 600 gallons per pulse, the totalizer will pulse once every minute.
AX2200/AX2300 Instruction Manual Chapter 3 Operation Totalizer #2 Menu Run Mode ENTER Password ENTER Totalizer Menu Use keys to access menus Totaling Inactive Mass Volume Energy Reset Total ? YES or NO Use the Totalizer #2 to Monitor Flow or Energy. Note that Totalizer #2 does not operate a relay, it is for monitoring only. CM2-AX2000-2001 (Rev.
Chapter 3 Operation AX2200/AX2300 Instruction Manual Energy Menu – For EMS Energy Meters Only Configuration: There are several possibilities regarding the measurement of water or steam energy given the location of the meter and the use of a second RTD.
AX2200/AX2300 Instruction Manual Chapter 3 Operation Fluid Menu Use the Fluid Menu to configure the flow meter for use with common gases, liquids and steam. Your flow meter is pre-programmed at the factory for your application’s process fluid. Reference Richard W. Miller, Flow Measurement Engineering Handbook (Third Edition, 1996), page 2-75 for definition and use of the Goyal-Doraiswamy equation and page 2-76 for the definition and use of the API 2540 equation.
Chapter 3 Operation AX2200/AX2300 Instruction Manual Units Menu Use the Units Menu to configure the flow meter with the desired units of measurement. (These are global settings and determine what appears on all screens. 3-14 CM2-AX2000-2001 (Rev.
AX2200/AX2300 Instruction Manual Chapter 3 Operation Time & Date Menu Use the Time and Date Menu to enter the correct time and date into the flow meter’s memory. The parameters are used in the Run Mode and the alarm and system log files. Note: Time is displayed in AM/PM format, but military format is used to set the time. For example, 1:00 PM is entered as 13:00:00 in the Set Time menu. Example for Setting the Time How to set the time to 12:00:00.
Chapter 3 Operation AX2200/AX2300 Instruction Manual Diagnostics Menu Run Mode ENTER Password ENTER Diagnositcs Menu Simulate Vortex Frequency (Hz) Sim Vor Freq xxx * Simulate Temperature Sim Temp 1, 2 xxx * Simulate Pressure Sim Pressure xxx Highest Recorded Velocity Highest Velocity xxx Highest Recorded Temperature Highest Temp 1, 2 xxx Highest Recorded Pressure Highest Pressure xxx Highest Electronics Temperature Highest Int Temp xx Lowest Electronics Temperature Lowest Int Temp xx U
AX2200/AX2300 Instruction Manual Chapter 3 Operation If the meter display indicates a temperature or pressure fault, a substitute value can be entered to allow flow calculations to continue at a fixed value until the source of the fault is identified and corrected. The units of measure of the displayed values are the same as the units configured for the flow meter. Calibration Menu The Calibration Menu contains the calibration coefficients for the flow meter.
Chapter 3 Operation AX2200/AX2300 Instruction Manual Password Menu Use the Password Menu to set or change the system password. The factory-set password is 1234. 3-18 CM2-AX2000-2001 (Rev.
AX2200/AX2300 Instruction Manual Chapter 4 Serial Communications Chapter 4 Serial Communications HART Communications The HART Communications Protocol (Highway Addressable Remote Transducer Protocol) is a bidirectional digital serial communications protocol. The HART signal is based on the Bell 202 standard and is superimposed on 4-20 mA Output 1. Peer-to-peer (analog / digital) and multi-drop (digital only) modes are supported.
Chapter 4 Serial Communications AX2200/AX2300 Instruction Manual DC Powered Meter Wiring Figure 4-2.DC Powered Meter Wiring (HART) AC Powered Meter Wiring Vortex Meter Current Meter Field Connection for Communicator 4-2 R load, 250 ohm minimum Remote Connection for Communicator CM2-AX2000-2001 (Rev.
AX2200/AX2300 Instruction Manual Chapter 4 Serial Communications Figure 4-3.AC Powered Meter Wiring (HART) CM2-AX2000-2001 (Rev.
Chapter 4 Serial Communications AX2200/AX2300 Instruction Manual HART Commands with the DD Menu Online Menu 1 Device Setup 1 Display Unit 1 Mass flo unit 2 Vol unit 3 Temp unit 4 Energy flo unit 5 Line press unit 6 Dens unit 7 Totalizer units 8 Std & Norm Cond 2 Analog Output 3 External Loop 4 Meter Display 5 Alarm Setup 6 Totalizer To Analog Output Menu 1 External Input 2 Set Ext. 4 mA 3 Set Ext. 20 mA 1 Disp Cycle 2 Disp Digits 3 Disp Damping 4 Init Displ.
AX2200/AX2300 Instruction Manual Chapter 4 Serial Communications HART Commands with the DD Menu Continued Analog Output Menu From Online Menu 1 Fix Analog Output 2 Trim Analog Output 3 Configure AO1 4 PV is 5 PV AO1 Out 6 PV % rnge 7 Configure AO2 8 SV is 9 SV AO2 Out SV % rnge Configure AO3 TV is TV AO TV % rnge Configure AO4 QV is QV AO QV % rnge 1 PV is 2 PV AO1 Out 3 PV 4 PV % rnge 5 Apply values 6 PV Rnge unit 7 PV LRV 8 PV URV 9 PV AO1 Lo end pt PV AO1 Hi end pt PV AO1 Added damp 1 SV is 2 SV AO
Chapter 4 Serial Communications AX2200/AX2300 Instruction Manual HART Commands with the DD Menu Continued 4-6 CM2-AX2000-2001 (Rev.
AX2200/AX2300 Instruction Manual Chapter 4 Serial Communications HART Commands with the DD Menu Continued CM2-AX2000-2001 (Rev.
Chapter 4 Serial Communications AX2200/AX2300 Instruction Manual HART Commands with the DD Menu Continued Sensor Cal Menu To Calibration Review Menu From Online Menu 1 Calibration Review 2 Vortex Sensor 3 Vortex Cal 4 Press Sensor 5 Press Cal 6 Temp Sensor 7 Temp 1 & 2 Cal 8 Temp 2 Sensor 9 Cal.
AX2200/AX2300 Instruction Manual Chapter 4 Serial Communications HART Commands with Generic DD Menu Online Menu 1 Device Setup 2 PV 3 PV AO 1 Process Variables 2 Diag/Service 3 Basic Setup 1 Snsr 2 AI % Rnge 3 AO1 1 Test Device 2 Loop Test 3 Calibration 4 D/A Trim 1 Tag 2 PV unit 3 Range Values 4 Device Information 5 PV Xfer fnctn 6 PV Damp 1 4 mA 2 20 mA 3 Other 4 End 1 Apply Values 2 Enter Values 1 PV LRV 2 PV URV 3 PV LSL 4 PV USL 1 Distributor 2 Model 3 Dev id 4 Tag 5 Date 6 Write Protect 7 De
Chapter 4 Serial Communications AX2200/AX2300 Instruction Manual Fast Key Sequence Use password 16363.
AX2200/AX2300 Instruction Manual Chapter 4 Serial Communications Sequence Description Access Notes 1,4,3,1,3,1 1,4,3,1,3,2 1,4,3,1,3,3 1,4,3,1,3,4 1,4,3,1,4 1,4,3,1,5 1,4,3,2,1 1,4,3,2,2 1,4,3,2,3 1,4,3,2,4 1,4,4,1 1,4,4,2 1,4,4,3 1,4,4,4 1,4,4,5 1,4,4,6 1,4,4,7 1,4,4,8 1,4,4,9 1,4,4,menu 1,4,4,menu,1 1,4,4,menu,2 1,4,4,menu,3 1,5 2 3 4,1 4,2 5,1 5,2 4 mA 20 mA Other End D/A trim Scaled D/A trim Poll addr Num req.
Chapter 4 Serial Communications AX2200/AX2300 Instruction Manual Modbus Communications Applicable Flow Meter Models Warning! Place controls in manual mode when making configuration changes to the vortex meter. azbil AX2000 Series Mass Flow Meters, Models AX2200 and AX2300 with Modbus communication protocol and firmware version 4.00.58 and above.
AX2200/AX2300 Instruction Manual Chapter 4 Serial Communications “RS-485 +” = “B” = “TxD+/RxD+” = “Non-Inverting pin” “RS-485 GND” = “GND” = “G” = “SC” = “Reference” CM2-AX2000-2001 (Rev.
Chapter 4 Serial Communications AX2200/AX2300 Instruction Manual Menu Items The following menu items are in the Output Menu and allow selection and control of the Modbus communication protocol. Address When the Modbus protocol is selected, the Modbus address is equal to the user programmable device address if it is in the range 1…247, in accordance with the Modbus specification. If the device address is zero or is greater than 247, then the Modbus address is internally set to 1.
AX2200/AX2300 Instruction Manual Chapter 4 Serial Communications The following four selections are available in this menu; when selecting an item, the protocol is changed immediately without having to press the Enter key.
Chapter 4 Serial Communications AX2200/AX2300 Instruction Manual Each range of register numbers maps to a unique range of addresses that are determined by the function code and the register number. The address is equal to the least significant four digits of the register number minus one, as shown in the following table.
AX2200/AX2300 Instruction Manual Registers 65100-65101 30525-30526 32037-32042 30009-30010 30007-30008 30005-30006 30001-30002 30029-30030 30015-30016 30013-30014 30031-30032 30025-30026 34532 30085-30086 30027-30028 Variable Chapter 4 Serial Communications Data type Serial number unsigned long Totalizer unsigned long Totalizer units string Mass flow float Volume flow float Pressure float Temperature float Velocity float Density float Viscosity float Reynolds number float Vortex frequency float Gain ch
Chapter 4 Serial Communications AX2200/AX2300 Instruction Manual reading a single 16-bit register, then the value will likely be invalid. The floating point registers with values in display units are scaled to the same units as are displayed, but are instantaneous values that are not smoothed. If display smoothing is enabled (non-zero value entered in the Display TC item in the Display Menu), then the register values will not agree exactly with the displayed values.
AX2200/AX2300 Instruction Manual Chapter 4 Serial Communications Control Register Definitions The only writeable registers in this implementation are the Reset Exception Status, Reset Meter and Reset Totalizer functions, which are implemented as ”coils” which may be written with the Write Single Coil command (function code 05) to address 7 through 9, respectively, (register #00008 through #00010).
Chapter 4 Serial Communications AX2200/AX2300 Instruction Manual Exception Response Message Format Examples Read the exception status byte from the device with address 1: 01 07 41 E2 01 Device address 07 Function code, 07 = read exception status 41 E2 CRC A typical response from the device is as follows: 01 07 03 62 31 01 07 03 62 Device address Function code Exception status byte 31 CRC Request the first 12 registers from device with address 1: 01 04 00 00 00 0C F0 0F 01 04 00 00 F0 Device address F
AX2200/AX2300 Instruction Manual 01 04 00 00 F0 Chapter 4 Serial Communications Device address Function code 4 = read input register 00 Starting address 50 Number of registers = 80 36 CRC results in an error response as follows: 01 84 02 C2 C1 01 Device address 84 Function code with most significant bit set indicates error response 02 Exception code 2 = invalid data address C2 C1 CRC Request the state all three alarms: 01 02 00 00 00 03 38 0B 01 02 00 00 38 Device address Function code 2 = read discret
Chapter 4 Serial Communications AX2200/AX2300 Instruction Manual The unit responds with an identical message to that transmitted, and the totalizer is reset. If the “coil” is turned off as in the following message, the response is also identical to the transmitted message, but the totalizer is not affected. 01 05 00 09 00 00 1D C8 01 05 00 00 1D 4-22 Device address Function code 5 = write single coil 09 Coil address = 9 00 Data to “turn off coil” does not reset totalizer C8 CRC CM2-AX2000-2001 (Rev.
AX2200/AX2300 Instruction Manual Chapter 4 Serial Communications BACnet MS/TP Communications 1. BACnet MS/TP Description The BACnet Master-Slave/Token-Passing (MSTP) driver implements a data link protocol that uses the services of the RS485 physical layer. The MS/TP bus is based on BACnet standard protocol SSPC-135, Clause 9. BACnet MS/TP protocol is a peerto-peer, multiple master protocols based on token passing.
Chapter 4 Serial Communications AX2200/AX2300 Instruction Manual 12. Change the required settings and press Exit & Enter buttons to save the configuration 13. Do steps from b to g, and change the comm. Type as Hart. 14. Reboot the device by power off and on. Note: a. IUT support 9600, 19200, 38400 baud rates b. MAC address range is 0‐127 3. Supported BACnet Objects A BACnet object represents physical or virtual equipment information, as a digital input or parameters.
AX2200/AX2300 Instruction Manual Protocol_Object_Types_Supported Object_List Max_ADPU_Length_Accepted Segmentation_Supported ADPU_Timeout Number_Of_ADPU_Retries Max_Masters Max_Info_Frames Device_Address_Binding Database_Revision Status_Flags Event_State Reliability Out_Of_Service Units Polarity Priority_Array Relinquish_Default Status_Flag Present_Value Inactive_Text Active_Text (W) – Writable Property.
Chapter 4 Serial Communications protocol‐object‐types‐ supported object‐list AX2200/AX2300 Instruction Manual {T,F,F,T,F,T,F,F,T,F,F,F,F,F,F,F,F,F,F,F,F,F,F,F,F} max‐apdu‐length‐ accepted segmentation‐supported apdu‐timeout number‐of‐APDU‐retries max‐master max‐info‐frames device‐address‐binding database‐revision {(analog‐input,1),(analog‐input,2),(analog‐input,3),(analog‐input,4), (analog‐input,5), (analog‐input,6),(analog‐input,7),(analog‐input,8) (analog‐input,9),(analog‐input,10), (analog‐input,11)
AX2200/AX2300 Instruction Manual Chapter 4 Serial Communications pounds‐mass‐per‐hour, tons‐per‐hour, grams‐per‐second , grams‐per‐minute 3 Temperature 1 4 Temperature 2 5 Pressure 6 7 Density Energy Flow 8 Totalizer 1 & Totalizer 2 CM2-AX2000-2001 (Rev.
Chapter 4 Serial Communications 10 11 12 13 14 15 16 17 18 19 4-28 StatusRegister Channel 1 (4‐20mA) Channel 2 (4‐20mA) Channel 3 (4‐20mA) Scaled Freq Flow Velocity Viscosity Frequency VorTex Amp FilterSetting AX2200/AX2300 Instruction Manual minute, liters‐per‐minute, liters‐per‐second, liters‐per‐hour, cubic‐meters‐per‐second, cubic‐meters‐per‐minute, cubic‐meters‐per‐hour If Totalizer selection for Energy measure – Kilowatts, Horsepower, btus‐per‐hour,, kilo‐btus‐per‐hour, megawatts NO UNITS milliam
AX2200/AX2300 Instruction Manual Chapter 4 Serial Communications 3.3. Binary Input Object: Vortex Mass Flow Meters Binary Input type objects are described in the below Table. Object Instance 1 2 3 Object Name Alarm1 Alarm2 Alarm3 4 External Description The status of the three alarms may be monitored via the Modbus command. The value returned indicates the state of the alarm, and will be 1 only if the alarm is enabled and active.
Chapter 4 Serial Communications AX2200/AX2300 Instruction Manual 4. ANNEX ‐ BACnet PROTOCOL IMPLEMENTATION CONFORMANCE STATEMENT Date: 19-April-2012 Vendor Name: Azbil Instruments Product Name: AX2200 multivariable flow-meter Product Model Number: AX2200/AX2300 VT/VTP Applications Software Version: 1.
AX2200/AX2300 Instruction Manual Chapter 4 Serial Communications Services Supported Read Property Write Property Read Property Multiple Write Property Multiple Who‐Is I‐Am Who‐Has I‐Have Device Communication Control Segmentation Capability: Able to transmit segmented messages Able to receive segmented messages CM2-AX2000-2001 (Rev.
Chapter 4 Serial Communications AX2200/AX2300 Instruction Manual Standard Object Types Supported: Standard Object Types Supported Dynamically Dynamically Additional Range Creatable Delete‐able Writable Properties Restrictions Analog Input (AI) No No None None Binary Input (BV) No No None None Binary Value No No None None Device No No None None Object Type Standard Object Types Supported Writable Properties Object Type Properties Analog Input (AI) Present Value Out‐Of‐Service Binary Input (BV) Present V
AX2200/AX2300 Instruction Manual Chapter 4 Serial Communications Object List: Properties of Analog Input/Value Objects Type ID Name AI1 AI2 AI3 AI4 AI5 AI6 AI7 AI8 AI9 AI10 Volume Flow Mass Flow Temperature 1 Temperature 2 Pressure Density Energy Flow Totalizer 1 Totalizer 2 StatusRegister Channel 1 (420mA) Channel 2 (420mA) Channel 3 (420mA) Scaled Freq Flow Velocity Viscosity Frequency VorTex Amp FilterSetting AI11 AI12 AI13 AI14 AI15 AI16 AI17 AI18 AI19 Present Value ? ? ? ? ? ? ? ? ? ? Status Fl
Chapter 4 Serial Communications AX2200/AX2300 Instruction Manual Properties of Analog Input/Value Objects Type ID Name BV1 Reset Present Value ? Status Flags F,F,F,F Event State Normal Out of Service False out-of-service False Data Link Layer Options: BACnet IP, (Annex J) BACnet IP, (Annex J), Foreign Device ISO 8802-3, Ethernet (Clause 7) ANSI/ATA 878.1, 2.5 Mb. ARCNET (Clause 8) ANSI/ATA 878.
AX2200/AX2300 Instruction Manual Chapter 4 Serial Communications Multiple Application‐Specific Keys: Supports encryption (NS‐ED BVBB) Key Server (NS‐KS BVBB) Character Sets Supported: Indicating support for multiple character sets does not imply that they can all be supported simultaneously. ANSI X3.
Chapter 4 Serial Communications AX2200/AX2300 Instruction Manual 5. Acronyms and Definitions Item APDU BACnet MS/TP BIBB BV BI AI RP WP RPM WPM DDB DOB DCC 4-36 Description Application Protocol Data Unit Building Automation and Control Network- Data communication protocol Master-Slave Token passing(a twisted pair RS485 network created by BACnet) BACnet Interoperability Building Block (Specific individual function blocks for data exchange between interoperable devices).
AX2200/AX2300 Instruction Manual Chapter 5 Troubleshooting & Repair Chapter 5 Troubleshooting and Repair Warning! Before attempting any flow meter repair, verify that the line is not pressurized. Always remove main power before disassembling any part of the mass flow meter. Hidden Diagnostics Menus The menus shown on the following page can be accessed using the password 16363, then moving to the display that reads “Diagnostics Menu” and pressing ENTER (rather than one of the arrow keys).
Chapter 5 Troubleshooting & Repair AX2200/AX2300 Instruction Manual Hidden Diagnostics Menus 5-2 CM2-AX2000-2001 (Rev.
AX2200/AX2300 Instruction Manual Chapter 5 Troubleshooting & Repair Level One Hidden Diagnostics Values f = Vortex shedding frequency (Hz). fi = Adaptive filter – should be approximately 25% higher than the vortex shedding frequency, this is a low-pass filter. If the meter is using the Filter Control (see below) in the manual mode, fi will be displayed as fm. G = Gain (applied to vortex signal amplitude). Gain defaults to 1.
Chapter 5 Troubleshooting & Repair AX2200/AX2300 Instruction Manual Lvl = Threshold level. If the Low Flow Cutoff in the calibration menu is set above this value, the meter will read zero flow. The Lvl level can be checked at no flow. At no flow, the Lvl must be below the Low Flow Cutoff setting or the meter will have an output at no flow. Adj. Flilter = Adjustable filter. Displays the filtering in decibels. Normally reads zero.
AX2200/AX2300 Instruction Manual Chapter 5 Troubleshooting & Repair 4-20(2) FScale = Analog counts to cal. full scale on analog output 2. 4-20(3) Zero = Analog counts to calibrate zero on analog output 3. 4-20(3) FScale = Analog counts to cal. full scale on analog output 3. Ext. 4 mA Cal. = Enter 0 for auto calibration or enter factory supplied A/D counts. Note: You must connect a known 4.00 mA input if you are going to calibrate the unit. Ext. 20 mA Cal.
Chapter 5 Troubleshooting & Repair 5-6 AX2200/AX2300 Instruction Manual Factory Defaults = Reset factory defaults. If you change this to Yes and press Enter, all the factory configuration is lost and you must reconfigure the entire program. Consult the factory before performing this process, it is required only in very rare cases. Meter Type = Insertion (AX2300) or Inline (AX2200) meter. Config Code = Factory use only. Test Pulse Out = Force totalizer pulse.
AX2200/AX2300 Instruction Manual Chapter 5 Troubleshooting & Repair Init Displ. (sec) = Enter a value in seconds to initialize the display every xxx seconds. Enter a value of 0 to disable initializing the display. Analog Output Calibration To check the 4–20 mA circuit, connect a DVM in series with the output loop. Select zero or full scale (from the second level of the hidden diagnostics) and then actuate the enter key twice. This action will cause the meter to output its 4 mA or 20 mA condition.
Chapter 5 Troubleshooting & Repair AX2200/AX2300 Instruction Manual Troubleshooting the Flow Meter Warning! Before attempting any flow meter repair, verify that the line is not pressurized. Always remove main power before disassembling any part of the mass flow meter. Use hazardous area precautions if applicable. Static sensitive electronics - use electro-static discharge precautions.
AX2200/AX2300 Instruction Manual Chapter 5 Troubleshooting & Repair Record values - Hidden Diagnostics Menu continued: With Flow With No Flow (if possible) Ck = Lvl = Adj. Filter = Iso. Power Volts = Record the following values from the Calibration Menu. Meter Size / Pipe ID = Meter Factor = Vortex Coef Ck = Low Flow Cutoff = Serial Number = Determine the Fault Symptom: Output at no Flow 1. The low flow cutoff is set too low.
Chapter 5 Troubleshooting & Repair AX2200/AX2300 Instruction Manual 3. The meter may be reacting to actual changes in the flow stream. The output can be smoothed using a time constant. The displayed values can be smoothed using the time constant in the Display Menu. The analog outputs can be smoothed using the time constant in the Output Menu. A time constant of 1 will result in the change in value reaching 63% of its final value in one second. A time constant of 4 is 22%, 10 is 9.5% and 50 is 1.
AX2200/AX2300 Instruction Manual Chapter 5 Troubleshooting & Repair PRESSURE VORTEX TEMPERATURE Figure 5-1.Electronics Stack Sensor Connections VORTEX PRESSURE TEMPERATURE Figure 5-2.Remote Feed Through Board Sensor Connections Symptom: No Output 1. For remote mounted electronics, carefully check all the wiring connections in the remote mount junction box. There are 18 connections that must be correct, verify each color (black and red), shield, and wire number. 2.
Chapter 5 Troubleshooting & Repair AX2200/AX2300 Instruction Manual OUTSIDE INSIDE OUTSIDE Figure 5-3.Vortex Sensor Connector With the sensor still disconnected, go to the first level of the hidden diagnostics and display the vortex shedding frequency, f. Hold a finger on the three exposed pins on the analog board. The meter should read electrical noise, 60 Hz for example. If all readings are correct, re-install vortex sensor wires. 4.
AX2200/AX2300 Instruction Manual Chapter 5 Troubleshooting & Repair OUTSIDE OUTSIDE Figure 5-4.Temperature Sensor Connector 4. Consult factory with findings Symptom: Meter Displays Pressure Fault 1. For remote mounted electronics, carefully check all the wiring connections in the remote mount junction box. There are 18 connections that must be correct, verify each color (black and red), shield, and wire number. 2.
Chapter 5 Troubleshooting & Repair AX2200/AX2300 Instruction Manual Electronics Assembly Replacement (All Meters) The electronics boards are electrostatically sensitive. Wear a grounding wrist strap and make sure to observe proper handling precautions required for static-sensitive components. 1. Turn off power to the unit. Warning! Before attempting any flow meter repair, verify that the line is not pressurized. Always remove main power before disassembling any part of the mass flow meter. 2.
AX2200/AX2300 Instruction Manual Chapter 5 Troubleshooting & Repair Pressure Sensor Replacement (AX2200 Models Only) 1. For local mounted electronics, remove the electronics stack as previously described. For remote mount electronics, remove all wires and sensor connectors from the remote feed through board in the junction box at the meter. 2. Loosen the three set screws at the center of the adapter between the meter and the enclosure. 3.
Chapter 5 Troubleshooting & Repair AX2200/AX2300 Instruction Manual 5-16 CM2-AX2000-2001 (Rev.
Series AX2200/AX2300 Instruction Manual Appendix A Specifications Appendix A Product Specifications Accuracy Process Variables AX2200 Model In-Line Meters AX2300 Model Insertion Me(1) ters Liquids Gas & Steam Liquids Gas & Steam Mass Flow Rate ±1% of rate over a 30:1 (3) range ±1.5% of (2) rate over a (3) 30:1 range ±1.5% of rate over a 30:1 (3) range ±2% of (2) rate over a (3) 30:1 range Volumetric Flow Rate ±0.7% of rate over a 30:1 (3) range ±1% of rate over a 30:1 (3) range ±1.
Appendix A Specifications Pressure 0 psig 100 psig 200 psig 300 psig 400 psig 500 psig AX2200/AX2300 Instruction Manual 0.5 1.8 18 5 138 7 258 8 380 10 502 11 624 Typical Air Minimum and Maximum Flow Rates (SCFM) o Air at 70 F Nominal Pipe Size (in) 0.75 1 1.
Series AX2200/AX2300 Instruction Manual Pressure 5 psig 100 psig 200 psig 300 psig 400 psig 500 psig Pressure 0 barg 5 barg 10 barg 15 barg 20 barg 30 barg Appendix A Specifications Typical Saturated Steam Minimum and Maximum Flow Rates (lb/hr) Nominal Pipe Size (in) 0.5 0.75 1 1.5 2 3 4 6 6.
Appendix A Specifications AX2200/AX2300 Instruction Manual AX2300 Pressure Ratings Probe Seal Compression Fitting Packing Gland Packing Gland with Removable Retractor Packing Gland with Permanent Retractor Process Connection Material Rating Ordering Code 2-inch MNPT 316L SS ANSI 600 lb CNPT 2-inch 150 lb flange, DN50 PN16 316L SS ANSI 150 lb, PN16 C150, C16 2-inch 300 lb flange, DN50 PN40 316L SS ANSI 300 lb, PN40 C300, C40 2-inch 600 lb flange, DN50 PN64 316L SS ANSI 600 lb, PN64 C
Series AX2200/AX2300 Instruction Manual Appendix A Specifications User is responsible for the provision of an external Disconnect Means (and Over-Current Protection) for the equipment (both AC and DC models). Display Alphanumeric 2 x 16 LCD digital display. Six push-button switches (up, down, right, left, enter, exit) operable through explosion-proof window using hand-held magnet. Viewing at 90-degree mounting intervals.
Appendix A Specifications A-6 AX2200/AX2300 Instruction Manual Mounting Position AX2200 In-Line Flow Meter: No effect. AX2300 Insertion Flow Meter: Meter must be perpendicular within ± 5° of the pipe centerline. Certifications Material Certificate – US Mill certs on all wetted parts Pressure Test Certificate Certificate of Conformance NACE Certification (MR0175) Oxygen Cleaning (CGA G-4. CM2-AX2000-2001 (Rev.
Series AX2200/AX2300 Instruction Manual CM2-AX2000-2001 (Rev.
Appendix A Specifications AX2200/AX2300 Instruction Manual A-8 CM2-AX2000-2001 (Rev.
AX2200/AX2300 Instruction Manual Appendix B Approvals Appendix B Approvals FM / FMC Approval Class I, Division 1, Groups B, C, & D, Class II/III, Division 1, Groups E, F, & G Type 4X and IP66, T6 Ta = -40 to 60°C ATEX-IECEx Specifications / Approval EN IEC 60079-0 (2006) Electrical Apparatus for explosive gas atmospheres General Requirements EN IEC 60079-1 (2007) Electrical Apparatus for explosive gas atmospheres Flameproof enclosures “d” EN IEC 61241-0 (2006) EN IEC 61241-1 (2004) Directive 94/9/EC (199
Appendix B Approvals AX2200/AX2300 Instruction Manual EC- Declaration of Conformity We VorTek Instruments 8475 West I‐25 Frontage Rd. Longmont, CO 80504 USA hereby declare in our sole responsibility, that the product AX2000 Series Models AX2200, AX2300 Multiparameter Vortex Flowmeters which is the subject of this declaration, is in conformity with the following standard(s), or normative documents. Terms of the directive Title and/or No.
AX2200/AX2300 Instruction Manual Appendix C Flow Meter Calculations Appendix C Flow Meter Calculations In-Line Flow Meter Calculations Volume Flow Rate QV f K Mass Flow Rate QM QV Flowing Velocity Vf QV A Where: A = Cross sectional area of the pipe (ft^2) f = Vortex shedding frequency (pulses / sec) K = Meter factor corrected for thermal expansion (pulses / ft^3) QM = Mass flow rate (lbm / sec) Qv = Volume flow rate (ft^3 / sec) Vf = Flowing velocity (ft / sec) = Density (lbm / ft^3) CM
Appendix C Flow Meter Calculations AX2200/AX2300 Instruction Manual Insertion Flow Meter Calculations Flowing Velocity Vf f Kc Volume Flow Rate QV V f A Mass Flow Rate QM V f A Where: A = Cross sectional area of the pipe (ft^2) f = Vortex shedding frequency (pulses / sec) Kc = Meter factor corrected for Reynolds Number (pulses / ft) Qv = Volume flow rate (ft^3 / sec) QM = Mass flow rate (lbm / sec) Vf = Flowing velocity (ft / sec) = Density (lbm / ft^3) C-2 CM2-AX2000-2001 (Rev.
AX2200/AX2300 Instruction Manual Appendix C Flow Meter Calculations Fluid Calculations Calculations for Steam T & P When “Steam T & P” is selected in the “Real Gas” selection of the Fluid Menu, the calculations are based on the equations below. Density The density of steam is calculated from the formula given by Keenan and Keys. The given equation is for the volume of the steam. v 4.555.04 T B p B B0 B0 g1 ( ) p B0 g 2 ( ) 3 p 3 B0 g 3 ( ) 12 p12 2 4 B0 1.89 2641.
Appendix C Flow Meter Calculations AX2200/AX2300 Instruction Manual Calculations for Gas (“Real Gas” and “Other Gas”) Use this formula to determine the settings for “Real Gas; Gas” selections and “Other Gas” selections entered in the Fluid Menu. The calculations for gas were taken from Richard W. Miller, Flow Measurement Engineering Handbook (Third Edition, 1996).
AX2200/AX2300 Instruction Manual Appendix C Flow Meter Calculations Calculations for Liquid Use this formula to determine the settings for “Goyal-Dorais” selections and “Other Liquid” selections entered in the Fluid Menu. The liquid calculations were taken from Richard W. Miller, Flow Measurement Engineering Handbook (Third Edition, 1996). Density The liquid density is found using the Goyal-Doraiswamy Equation.
Appendix C Flow Meter Calculations AX2200/AX2300 Instruction Manual C-6 CM2-AX2000-2001 (Rev.
AX2200/AX2300 Instruction Manual Appendix D Glossary Appendix D Glossary A B C D A Cross sectional area. ACFM Actual Cubic Feet Per Minute (volumetric flow rate). ASME American Society of Mechanical Engineers. Bluff Body A non-streamlined body placed into a flow stream to create vortices. Also called a Shedder Bar. BTU British Thermal Unit, an energy measurement. Cenelec European Electrical Code.
Appendix D Glossary AX2200/AX2300 Instruction Manual I J K L In-Line Flow Meter A flow meter which includes a short section of piping which is put in-line with the user’s piping. Insertion Flow Meter A flow meter which is inserted into a hole in the user’s pipeline. Joule A unit of energy equal to one watt for one second. Also equal to a Newton-meter. LCD Liquid crystal display. M N O P D-2 m Mass flow rate. mA Milli-amp, one thousandth of an ampere of current.
AX2200/AX2300 Instruction Manual PV Appendix D Glossary Liquid vapor pressure at flowing conditions (psia or bar absolute). Q R S T Q Flow rate, usually volumetric. Rangeability Highest measurable flow rate divided by the lowest measurable flow rate. Reynolds Number or Re A dimensionless number equal to the density of a fluid times the velocity of the fluid times the diameter of the fluid channel, divided by the fluid viscosity (i.e., Re = VD/).
Appendix D Glossary AX2200/AX2300 Instruction Manual D-4 CM2-AX2000-2001 (Rev.
Azbil Corporation