User Manual

www.bk precision.com
Tel.: 714.921.9095
DC Current
Ranges 660 uA, 6600 uA, 66 mA, 400 mA, 10 A
Resolution 0.1 uA, 1 uA, 10 uA, 100 uA, 10 mA
A
ccuracy 660 uA to 400 mA ranges: ±(1.5 % rdg + 2 dgts)
1
0 A range: ±(3.0 % rdg + 3 dgts)
Input Protection 0.5 A/500 V & 10 A/600 V fast blow ceramic fuses
1
0A Input 10 A for 60 sec max. followed by a 10 minute cooling period
AC Current (True RMS) (50 Hz -500 Hz)
R
anges 660 uA, 6600 uA, 66 mA, 400 mA, 10 A
Resolution 0.1 uA, 1 uA, 10 uA, 100 uA, 10 mA
Accuracy 660 uA to 400 mA ranges: ±(2.0 % rdg + 10 dgts)
1
0 A range: ±(3.5 % rdg + 10 dgts)
C
rest Factor 3
Input Protection 0.5 A/500 V & 10 A/600 V fast blow ceramic fuses
10A Input 10 A for 60 sec max. followed by a 10 minute cooling period
Resis tance
R
anges 660 , 6.6 k, 66 k, 660 k, 6.6 M, 66 M
Resolution 0.1 , 1.0 , 10 , 100 , 1 k, 10 k
Accuracy 660 to 660 k ranges: ±(1.2 % rdg + 5 dgts)
6
.6 M range: ±(2.0 % rdg + 5 dgts)
6
6 M range: ±(3.5 % rdg + 5 dgts)
O
pen Circuit Voltage (typical) -0.45 VDC (-1.2 VDC on 660 range)
Overload Protection 500 VDC/ACrms
Capacitance
Ranges 6.6 nF , 66 nF, 660 nF, 6.6 uF, 66 uF, 660 uF, 6.6 mF, 66 mF
Resolution 1 pF, 10 pF, 100 pF, 1 nF, 10 nF, 100 nF, 1 uF, 10 uF
A
ccuracy 6.6 nF range: ±(3.0 % rdg + 30 dgts)
66 nF to 660 uF ranges: ±(3.0 % rdg + 5 dgts)
6.6 mF to 66 mF ranges: ±(5.0 % rdg + 20 dgts)
Overload Protection 500 VDC/ACrms
Fre qu ency
Ranges 660 Hz, 6.6 kHz, 66 kHz, 660 kHz, 6.6 MHz, 66 MHz
Resolution 0.1 Hz, 1 Hz, 10 Hz, 100 Hz, 1 kHz, 10 kHz
Accuracy ±(0.1 % rdg + 3 dgts)
Sensitivity 10 Hz to 6.6 MHz: >2.5 Vrms, 6.6 MHz to 66 MHz:
>2.5 Vrms <5 Vrms
Minimum Pulse Width >25 ns
Duty Cycle Limits >30 % and <70 %
Overload Protection 500 VDC/AC rms
Diode Test
Test Current 1.2 mA (typical)
Resolution 1 mV
Accuracy ±(3.0 % rdg + 3 dgts)
Audible Indication <0.03 V
Open Circuit Voltage 3.5 VDC (typical)
Overload Protection 500 VDC/ACrms
Con tinuity
Audible Indication Less than 30
Response Time 100 ms
Overload Protection 500 VDC/ACrms
General
Display 3 3/4 digit, 6600 count LCD
Polarity Automatic, positive implied, negative polarity indication
Overrange OL or -OL is displayed
Low Battery Indication Battery symbol will display when battery voltage drops below
operating levels
Measurement Rate 2 times per second
Auto Power Off Approximately 30 minutes
Operating Environment 0ºC to 50ºC at <70 % relative humidity
Storage Environment -20ºC to 60ºC at 0 to 80 % relative humidity
Accuracies Stated accuracy at 23ºC +5ºC at <75 % relative humidity
Temperature Coefficient 0.1 x (specified accuracy) per ºC. (ºC to 18ºC, 28ºC to 50ºC)
Power Single standard 9 volt battery, NEDA 1604, JIS 006P, IEC 6F22
Battery Life 150 hours typical
Dimensions 165 mm (H) x 78 mm (W) x 42.5 mm (D)
Weight Approximately 10 oz. (285 g) with battery and hanger
Auto Ranging True RMS Tool Kit
®
DMM
Model 2709B
Data Sheet
Technical data subject to change
© B&K Precision Corp. 2013
Page 1 of 2
The 2709B is a full featured auto ranging true RMS DMM that combines
performance, value and functionality. The 6600 count LCD has a green back light
that allows the meter to be used in dimly lit environments. The 2709B can
measure resistance values up to 66 M and capacitance to 66 mF. With the auto
power off feature you won’t have to worry about accidental battery consumption.
Value-packed features make this meter a must in every "Tool Kit
®
".
Features
Magnetic hanger One-handed operation
True RMS Large easy-to-read LCD with backlight
Auto/manual ranging Auto power off
DC voltage to 1000 V Measures up to 10 A (AC & DC)
AC voltage to 750 V Resistance to 66 M
Capacitance to 66 mF Frequency counter to 66 MHz
Diode test Continuity test
DC Volts
Ranges 660 mV, 6.6 V, 66 V, 660 V, 1000 V
Resolution 0.1 mV, 1 mV, 10 mV, 100 mV, 1 V
Accuracy ±(0.5 % rdg + 2 dgts)
Input Impedance 660 mV: >100 M; 6.6 V: 10 M; 66 V to 1000 V: 9.1 M
Overload Protection 1000 VDC or 750 VACrms
AC Volts (T rue RM S) (50 Hz -500 Hz)
Ranges 660 mv, 6.6 V, 66 V, 660 V, 750 V
Resolution 0.1 mV, 1 mV, 10 mV, 100 mV, 1 V
Accuracy 660 mV range at 50 Hz to 60 Hz ±(1.5 % rdg + 8 dgts)
6.6 V to 660 V ranges: ±(1.5 % rdg + 8 dgts)
750 V range: ±(2.0 % rdg + 8 dgts)
Crest Factor 3
Input Impedance 660 mV: >100 M; 6.6 V: 10 M; 66 V to 750 V: 9.1 M
Overload Protection 1000 VDC or 750 VACrms
Specifications
2709B
model
Supplied Accessories: User manual, one set of test leads, one spare fuse, and 9V battery (installed)
Three-Year Warranty
RUE RMS
T
Select between Auto or
Manual ranging
Aut o/ Manual Ranging

Summary of content (2 pages)