4.13.1.6 Point Rate
:SOURce:PRATe <point rate>
This command is not used. Point rate setting is available for arbitrary mode only. Please see section 4.13.4.1 for details.
4.13.1.7 AM modulation
The following sections control the AM modulation:
4.13.1.7.1 AM STATe
This command activates or deactivates AM modulation:
Arguments
Type: Boolean
Command Type: Setting or Query
Setting
Syntax: [: SOURce:]AM [:STATe]<ws>ON|1|OFF|0
Examples: : SOURce:AM :STAT ON
AM OFF
Query
Syntax: [:SOURce:]AM[:STATe]?
Response: 0|1
4.13.1.7.2 AM DEPTh
This command sets the AM modulation depth in %
Arguments
Type: Numeric
Units: none (implied %)
Range: 0 to 100
Rounding: to integer
Command Type: Setting or Query
Setting
Syntax: :SOURce:AM:DEPTh<ws><percent depth>
:SOURce:AM:DEPTh<ws>MINimum|MAXimum
Examples: AM:DEPTh 50
Query
Syntax: AM:DEPTh?[<ws>MINimum|MAXimum]
Response: NR3
4.13.1.7.3 AM SHAPe
This command selects the AM modulating waveform shape
Arguments
Type: Character
Options: SINusoid, TRIangle, SQUare
Command Type: Setting or Query
Setting
Syntax: [:SOURce:]AM:SHAPe<ws><SIN|TRI|SQU>
Examples: [:SOURce:]AM:SHPE SIN
AM:SHAPE TRI
Query
Syntax: [:SOURce:]AM:SHAPe?
Response: SIN|TRI|SQU