Owner manual

4080 series Function & Arbitrary/Function Generators Instruction Manual
42
FM:INTernal:FUNCtion {SINusoid|SQUare|TRIangle……}
Set the modulating waveform of FM using SIN, SQU, TRI UP, RAMP or DOWN_RAMP waveforms.
FM:INTernal:FUNCtion?
Read the current modulating waveform of FM. The returned data are SIN”, SQU”, TRI”,
UP_RAMP”, “DOWN_RAMP”.
FM:INTernal:FREQuency <frequency>
Set the frequency of modulating signal of FM. The modulating frequency should be 10KHz when
internal modulating source is selected.
FM:INTernal:FREQuency?
Read the frequency of current modulating signal of FM.
FM:SOURce {INTernal|EXTernal}
Set modulating source of FM signal as internal (INTernal) or external (EXTernal).
FM:SOURce?
Read the current modulating source of FM.
FM:STATe {ON|OFF}
Turn on FM mode (ON), turn off FM mode (OFF).
FM:STATe?
Check the “on/off” state of FM mode. The returned data are 0 (OFF) or 1 (ON). It returns to
fixed frequency mode in “off” state.
Burst Mode instructions
1. Set the carrier using APPLy instructions or FUNCtion, FREQuency, VOLTage, VOLTAGE:
OFFSet instruction.
2. Set number of burst using BM: NCYCles <#cycles> instruction.
3. Set start phase of burst using BM: PHASe <degrees> instruction.
4. Set time interval between two bursts using BM:INTernal: SPACe <time> instruction.
5. Set trigger source of burst using BM: SOURce {INTernal|EXTernal|SINGle} instruction.
6. To turn on burst use the BM: STATe ON instruction.
BM:NCYCles <#cycles>
Set waveform numbers of burst. MAX10000.
BM:NCYCles?
Check number of bursts.
BM:PHASe <degrees>
Set start phase of burst in 0360º in step of 0.1º.
BM:PHASe?
Check the start phase of burst.
BM:INTernal: SPACe <time>
Set time interval between two bursts.
BM:INTernal: SPACe?
Check time interval between two bursts.
BM:SOURce {INTernal|EXTernal|SINGle}
Set trigger source of burst as internal (INTernal), external (EXTernal) or single (SINGle).