CHAPTER 3. OPERATION 23
Parameter On-screen Label Description/Options
Trigger Source TRIG
Selects the trigger source.
Options: INT, MAN, EXT, BUS
INT (Internal) – Automatic continuous measurement
MAN (Manual) – Each measurement is made by pressing the
buon.
EXT (External) – Measurement is made upon receiving a trig-
ger signal from either the rear panel BNC or handler in-
terface terminals.
BUS (Bus) – Measurement is made upon receiving a trigger
command from a remote interface.
Measurement Delay DELAY
Sets a measurement delay time.
Valid Range: 0 ms - 60 seconds
Note: Values entered are in milliseconds (ms).
Measurement Voltage
Display
Vm
Enables/disables the measured voltage display in the measure-
ment display screen.
Options: OFF, ON
Deviation Primary
Measurement Display
DEV_A
Enables/disables the deviation measurement display of the pri-
mary measurement function.
Options: OFF, ABS, %
ABS - Primary measurement display will show the dierence
between measured value and REF_A value.
% - Same as ABS, but measurement will show percentage dif-
ference instead.
Voltage Range RNG_V
Configures the measurement DC voltage range.
Options: AUTO, HOLD, 300V (*60V), 30V (*6V)
*Model BA6010
AUTO - Autoranging.
HOLD - Locks the current voltage range.
300V(60V) - Selects the 300 V/60 V range.
30V(6V) - Selects the 30 V/6 V range.
Current Range RNG_R
Configures the measurement AC impedance range.
Options: AUTO, HOLD, ,
AUTO - Autoranging.
HOLD - Locks the current impedance range.
, - Selects between 30mΩ, 300mΩ, 3Ω, 30Ω, 300Ω, 3kΩ
range.
Measurement Averaging AVG
Sets the number of samples for measurement averaging.
Valid Range: 1 to 255
Measurement Frequency LINE
Selects the frequency of the measurement source.
Options: 50Hz, 60Hz
Select the same frequency as that of the AC power source to
minimize measurement error.