CHAPTER 3. OPERATION 30
Figure 3.20: BIN Setup Display
Parameter On-screen Label Description/Options
Measurement Mode Mode
Selects the measurement mode for the BIN function.
Options: %, ABS
% — The high and low limits used for testing is a tolerance per-
centage.
ABS — The high and low limits used for testing is an absolute
value of the selected parameter.
HIGHA[], LOWA[], HIGHB[], LOWB[] — labels display ei-
ther [X] units, where ‘X’ denotes the primary and secondary
measurement units when applicable (i.e. Ω, V, H, F, r(radians)),
or [%] with the limits defining a percentage.
Comparator State
Primary Measurement
COMPA
Enables or disables the primary measurement comparator
function. When disabled, primary measurements will not be
tested.
Options: OFF, ON
Comparator State
Secondary Measurement
COMPB
Enables or disables the secondary measurement comparator
function. When disabled, secondary measurements will not be
evaluated.
Options: OFF, ON
Nominal Value
Primary Measurement
NOMA
Sets the nominal value of the bin comparator function for
the primary measurement. This parameter only applies when
Mode is set to %.
Valid Range: Dependent on the range of the primary measure-
ment selected.
Nominal Value
Secondary Measurement
NOMB
Sets the nominal value of the bin comparator function for the
secondary measurement. This parameter only applies when
Mode is set to %.
Valid Range: Dependent on the range of the secondary mea-
surement selected.
High Limit
Primary Measurement
HIGHA
Sets the upper limit value for the comparator function that ap-
plies to the primary measurement.
Valid Range: -100 - 100 (%), -10000 - 10000 (ABS)