Parameter On-screen Label Description/Options
Secondary Measurement
Upper Limit
B Stop 1
Sets the upper limit value as a stop condition of the test. If the
secondary measured value exceeds this for 2 consecutive sam-
ples, the test will stop.
Valid Range: Dependent on the measurement range of the se-
lected secondary function.
Secondary Measurement
Lower Limit
B Stop 2
Sets the lower limit value as a stop condition of the test. If the
secondary measured value exceeds this for 2 consecutive sam-
ples, the test will stop.
Valid Range: Dependent on the measurement range of the se-
lected secondary function.
Table 3.12: Trace Setup Parameters
Figure 3.22: Trace Display
The Trace Display, Figure 3.22 shows the measurements and trace results on the screen. To show this display,
press the buon, then pressing the function key corresponding to TRACE DISP from the so menu.
3.10.2 Trace Display Tools
The trace display tools menu, Figure 3.23, is accessed from the Trace Display by pressing and select
TRACE DISP, and then pressing the key twice to select TOOLS.
The trace display tools menu has options to start the trace test and to control what to display on the screen. See
Table 3.13.
3.10.3 Start a Trace
Configure all the seings under Trace Setup, and then go to the Trace Display Tools menu.
Select the TRACE to display then select SCAN START.
While testing, the display plots the measured values.
To stop testing at any time, select SCAN STOP.