User Manual

CHAPTER 3. OPERATION 34
Figure 3.23: Trace Tools Display
Parameter On-screen Label Description/Options
Initiate or stop trace
test
SCAN START
(SCAN STOP)
Starts or stops the trace test.
Options: START, STOP
Trace Display TRACE
Selects the trace to display.
Valid Range: A, B, A+B
A The primary measurement trace.
B The secondary measurement trace.
Cursor Display CURSOR
Enables/disables the cursor. When enabled, the cursor can be
controlled using the arrow keys. The cursor is used to
view the primary and secondary measurements, and the times-
tamp of any measured points on the trace.
Maximum/Minimum
Value Display
Max-Min
Enables/disables the display of the maximum and minimum
values of both the primary and secondary measurement traces.
Options: ON, OFF
Table 3.13: Trace Display Tools Parameters
Press CURSOR to see the values of a measured point on the trace. Using the keys to move the
cursor along the trace.
The maximum and minimum values of the primary and secondary measurements of the trace may also be
displayed by pressing Max-Min to ON.
3.11 Statistical Measurement
The instrument can calculate statistical measurements, which are accessed by pressing and selecting
STATIS DISP from the so menu.
Before performing a statistical calculation, some parameters must be configured first. They are grouped together
on display the box highlighted in cyan like this, parameter . See Figure 3.26. Use the arrow keys to select each
parameter, and the function keys to make changes. Parameters requiring entering a numeric values are entered