714-921-9095 bkprecision.com 7
batteries to power the probe. Set the probe attenuation ratio
to 10x.
2. Short the + and - probe inputs together with the hook tips.
3. Turn on power to the oscilloscope. Leave both the scope
and the probe on for 30 minutes to stabilize.
4. Press [Default Setup] and [Auto] on the oscilloscope.
5. Press the channel 1 button, then press the Probe soft key
and set the attenuation to 10x.
6. Set the scope to DC coupled mode and the scope offset to 0
volts.
7. Set the oscilloscope to average mode (x16) or high-
resolution mode to reduce oscilloscope noise.
8. Using the offset adjustment tool (included), adjust the probe
offset voltage to 0 volts
4 Cleaning
This probe does not require any particular cleaning. If necessary, clean
the case with a cloth slightly moistened with soapy water
WARNING
Dry the probe thoroughly before attempting to make voltage
measurements.
CAUTION
Do not subject the probe to solvents or solvent fumes as these can
cause deterioration of the probe body and cables.