User Manual
85 
level 
PVSIMulation 
PV Simulation subsystem 
:CURVe 
[:NUMber] <NR1> 
Set the curve number n to write the 
parameters to 
:PARAMeter <NRf, NRf, 
     NRf, NRf 
     | ?> 
Set or return the parameters 
(Regulation type, Vmp, Pmp and 
Material Type) of the curve number n 
Regulation type: 
1 = EN50530 
2 = SANDIA 
3 = NBT32004 
Material Type: 
EN50530 – 1 = cSi ; 2 = TF 
SANDIA – 1 = TF ; 2 = SCMC ; 3 = HEC 
NBT32005 – 1 = cSi ; 2 = TF 
Example (SANDIA, HEC, 1000 Vmp, 
2000 Pmp): 
PVSIM:CURV:PARAM 1,1000,2000,3 
:SELEct <NR1 | ?> 
Set or return the selected curve to be 
running 
:POWer <NRf | ?> 
Set or return the simulated output 
power from 0 ~ 100% 
[:STATe] <Bool | ?> 
Set or return the PV simulation 
function state. 
:CONTrol:MODe <CC|CV|?> 
Set or return the control operation 
mode for PV simulation. 
[SOURce] 
Source subsystem 
:CURRent 
[:LEVel] <NRf | ?> 
Set or return output current level 
:PROTection 
[:STATe] <Bool | ?> 
Set or return over-current protection 