User Manual
93 
For example, if the OVP, OCP, and OPP ON/OFF switches are ON and the 
output is ON, the return value will be E40000. 
Bit7 
Bit6 
Bit5 
Bit4 
Bit3 
Bit2 
Bit1 
Bit0 
String 
Byte0 
1 
1 
1 
0 
0 
1 
0 
0 
E4 
Byte1 
0 
0 
0 
0 
0 
0 
0 
0 
00 
Byte2 
0 
0 
0 
0 
0 
0 
0 
0 
00 
5  Calibration 
B&K Precision recommends a calibration interval of one year to ensure 
that the power supply meets specifications. This instrument features 
closed-case calibration. To perform the calibration, the following 
equipment is required: 
  5 ½ digit digital multimeter (DMM); B&K Precision 5491B or 
equivalent 
The following calibration instructions may be used by authorized 
technicians or calibration personnel only. If you are not 
authorized, do not attempt to calibrate the instrument yourself, 
as it may damage the instrument and void the warranty. 
To enter the calibration procedure, follow the steps below: 
1.  Press the   button and press the   button until CAL is 
selected and blinking and press  . 