User Manual
94 
2.  Enter the password 13579, using the numeric keypad, and press 
the   key to confirm. The following display will be shown: 
Below is a list of calibration options: 
1.  Output voltage 
2.  Output current 
3.  Overvoltage protection 
4.  Overcurrent protection 
5.  External voltage programming 
6.  External current programming 
7.  Constant current of external voltage programming 
8.  Constant current of external current programming 
These options are only accessible in the calibration menu. The following 
calibration procedures assume the operator is in the CAL menu. 
5.1  Voltage Calibration  
Follow the steps below to perform the voltage calibration: 
1.  Connect the DMM to the output of the power supply as shown 
in the figure below: 