User Manual
95 
Figure 5.1 – Voltage Calibration Diagram 
2.  Set the DMM to a DC voltage measurement. Select the 
VOLTAGE parameter in the CAL menu and press  . The 
following will be displayed: 
The P: 1/4 signifies there are 4 points to be calibrated, and the 
current calibration point is 1. 
3.  Input the voltage reading from the DMM in the METER 
parameter by using the numeric keypad and press  . 
4.  Repeat this procedure by entering the meter readings for points 
2 – 4. 
5.2  Current Calibration  