User Manual
96 
Follow the steps below to perform the current calibration: 
1.  Connect the DMM to the output of the power supply as shown 
in the figure below: 
Figure 5.2 – Current Calibration Diagram 
2.  Set the DMM to a DC current measurement. Select the 
CURRENT parameter in the CAL menu and press  . The 
following will be displayed: 
The P: 1/4 signifies there are 4 points to be calibrated, and the 
current calibration point is 1. 
3.  Input the current reading from the DMM in the METER 
parameter by using the numeric keypad and press  . 