User Manual
102 
Figure 5.6 – CC Calibration of External Voltage Diagram 
2.  Set the DMM to a DC current measurement. Select the 
EXT_V_CC parameter in the CAL menu and press  . The 
following will be displayed: 
3.  Input the current reading from the DMM into the METER 
parameter by using the numeric keypad and press  . 
5.8  CC Calibration of External Current 
Follow the steps below to perform the calibration procedure for the 
constant current source of the external current programming by resistor 
mode: 
1.  Connect the DMM to the external analog control terminals of 
DB25 connector (+ on the DMM to Pins 10, - on the DMM to Pin 
22(GND)) on the rear of the power supply, as shown in the 
figure below: 