User Manual
14 
to the amount of current used for operation: 
AWG 
6 
8 
10 
12 
14 
16 
18 
20 
22 
Imax(A) 
75 
55 
40 
25 
20 
13 
10 
7 
5 
mΩ/meter 
1.3 
2.1 
3.3 
5.2 
8.3 
13.2 
21 
33.5 
52.8 
Table 1 – Wire Gauge Rating 
Before connecting wires to the output terminals, turn 
OFF the power supply to avoid damage to the 
instrument and the device under test (DUT). For safety, 
load wires must have a wire gauge size large enough to 
prevent overheating when the power supply operates at 
maximum short circuit output current. It will also 
prevent large voltage drops from resistances in the 
wires. 
SHOCK HAZARD:  
Hazardous voltages may exist at the outputs and the 
load connections when using a power supply with a 
rated output greater than 40V. To protect personnel 
against accidental contact with hazardous voltages, 
ensure that the load and its connections have no 
accessible live parts. Ensure that the load wiring 
insulation rating is greater than to the maximum output 
voltage of the power supply. 