User Manual
35 
2.  Use the     keys to select the limit setting value of VMAX, 
VMIN, IMAX, and IMIN. Press   to confirm selection. 
3.  Use the numerical keys to set the limit value and press   to 
confirm the setting. 
4.  Press   several times to exit the menu setting. 
Protection Settings 
Configure Over Voltage Protection (OVP) 
The PVS overvoltage protection utilizes a hardware comparator that 
quickly protects the instrument when the voltage presented at the 
output terminal exceeds the OVP setting voltage. 
Follow the steps below to set the OVP limit: 
1.  Press the   button and press the   button one time until 
CONFIG is blinking and press  . 
2.  Press   button one time until PROTECT is blinking and press 
. The display will then show OVP. 
3.  Use rotary to select the OVP ON or OFF followed by  . 