User Manual
36 
4.  Use numerical keys directly or use     keys with rotary 
adjustment followed by   to confirm the OVP value. 
5.  Press   several times to exit the menu setting. 
6.  If the OVP is on, the OVP LED indicator will be lit up. 
When OVP protection is tripped during operation, the output will turn 
off and the following OVP status message will display: 
To clear the trip status, press   once. 
Configure Over Current Protection (OCP) 
The PVS overcurrent protection utilizes a hardware comparator that 
quickly protects the instrument when the current presented at the 
output terminal exceeds the OCP setting current. 
Follow the steps below to set the OCP limit: 
1.  Press the   button and press the   button one time until 
CONFIG is blinking and press  . 
2.  Press   button one time until PROTECT is blinking and press 
. The display will then show OVP.  
3.  Press   button one time until OCP is displayed. 