User Manual
57 
following screen should be displayed. 
While in parallel connection mode, the output voltage of 
each power supply should be set to equal value. If the 
voltage value of each unit is not the same, the higher 
output voltage will feed back to the lower voltage unit 
and damage its internal components. 
 DO NOT connect multiple power supplies in series as it 
will cause damage or malfunction. 
When connecting in parallel mode, only the MASTER 
unit can be controlled by a computer for remote 
communication. There is also a decrease in speed and 
performance with each additional SLAVE units 
connected in parallel and linked together.  The voltage 
and current scan speed, which affects synchronization 
between the supplies, between the slave units and the 
master unit will increase based on the following 
formula: 
200 + 50 * (N – 1) (in milliseconds) 