User Manual
72 
2.  Press   button two times until RS-485 is blinking and press 
 to configure the settings for RS-485 remote 
communication. The following display will be shown: 
3.  Use     keys to select between each settings and press 
 to configure the corresponding setting. 
4.  To change the parameter settings, use the rotary dial. 
5.  Press   to save each setting and the display will return to the 
RS485 menu. 
6.  Below lists the options that can be changed for each setting: 
Baud rate: 9600, 19200, 38400, 57600 
Parity/Data bits: None/8 bits, Even/8 bits, Odd/8 bits, Even/7 
bits, Odd/7 bits 
Stop bit: 1, 2   
Address: 1 to 50 
Chain: On, Off 
Note: The default is 9600, None/8 bits, 1, Address = 1, Off.  
7.  For each power supply that you want to control, provide a 
different address. For example, if you have three power 
supplies to control, set the first supply to address 1, second 
supply to address 2, and third supply to address 3. If two or 
more supplies have the same address, the RS485 protocols will 
create communication errors. 