5.4  OCP Calibration  
The overcurrent protection (OCP) can be executed right after the 
current calibration or the user can select OCP in the CAL menu.  
Follow the steps below to perform the voltage calibration: 
1.  Short the output + and - terminals of the power supply as shown 
below: 
Figure 5.3 – OCP Calibration Diagram 
2.  Select the OCP parameter in the CAL menu and press  . The 
following will be displayed: 
3.  Press   to run the OCP calibration or press   to abort. The 
OCP calibration is an automatic process that will take about 20 
to 30 seconds. 
4.  After the calibration is finished, the following message will be 
displayed: 