4.  Use rotary to select the OCP ON or OFF followed by  . 
5.  Use numerical keys directly or use     keys with rotary 
adjustment followed by   to confirm the OCP value. 
6.  Press   several times to exit the menu setting. 
7.  If the OCP is on, the OCP LED indicator will be lit up. 
When OCP protection is tripped during operation, the output will turn 
off and the following OCP status message will display: 
To clear the trip status, press   once. 
Configure Overpower Protection (OPP) 
The PVS overpower protection continuously monitors the output power 
level. If it is greater than the OPP setting, it turns off the power supply 
output to protect the device under test. 
Follow the steps below to set the OPP limit: 
1.  Press the   button and press the   button one time until 
CONFIG is blinking and press  . 
2.  Press   button one time until PROTECT is blinking and press 
. The display will then show OVP.  
3.  Press   button two times until OPP is displayed. 