value by connecting a resistance value of 0-5 kΩ (0-5 V / 0-5 kΩ mode) 
or 0-10 kΩ (0-10 V / 0-10 kΩ mode) between Pin 9 and Pin 22 as shown 
below. 
Figure 3.5 - Analog Voltage Programming Diagram (Resistor Mode) 
Current Program 
This function is able to program the current output by connecting an 
external DC voltage (voltage mode) or an external resistor (resistor 
mode) to Pin 10. For this function to be enabled, the output control 
must be in analog mode. The external voltage range or resistor range 
used to control the full scale of the output voltage can be selected 
between 0-5 V / 0-5 kΩ and 0-10 V / 0-10 kΩ range. Follow the steps 
below to configure the voltage program. 
1.  Press the   button and press the   button one time until 
CONFIG is blinking and press  . 
2.  Press   button three times until EXTCTRL is blinking and 
press  .  
3.  Press   button two times to select CURRENT and Press  .  
4.  Use the rotary knob to select EXT-V (external voltage) or EXT-
R(external resistance) option and press  . To have front 
panel control, select FRONT. 
5.  Use the     keys to select PROG RNG (program range) 
and press  . 
0 – 5 kΩ 
or 
0 – 10 kΩ 