User manual
50 
Voltage Monitor 
This function is able to monitor the voltage output using Pin 11 and one 
of the ground pins (i.e. Pin 22), which can be connected to a digital 
voltage meter (DVM) or other voltage monitoring device, as shown 
below. The output control must be in analog mode to use this function. 
The monitoring of the output voltage range (which reflects 0 to full scale 
of the power supply’s output voltage) can be selected between 0-5 V 
and 0-10 V.  
Figure 3.8 - Analog Voltage Monitor Diagram 
Follow the steps below to configure the voltage program. 
1.  Press the   button and press the   button one time until 
CONFIG is blinking and press  . 
2.  Press   button three times until EXTCTRL is blinking and 
press  .  
3.  Press   button four times to select MON RNG and Press 
.  
4.  Now use the rotary knob to select 5V or 10V and press  . 
5.  Press   several times to exit the menu setting. 
Current Monitor 
This function is able to monitor the voltage output using Pin 24 and one 