2.3 Connection cable
Control unit and scanner are connected by a 8-wire PUR-cable:
 6 pin small circular connector to DIN 45322 at the control unit end.
 8 pin molded plug at the scanner end.
 Length 5 m, can be extended to a maximum length of approximately
25 m with extension cables.
Pin configuration (at the control unit end)
Small circular connector to DIN 45322, 6 pin
1 wh
white
2 br brown
3 gn green
4 ye yellow
5 gr gray
6 pi pink
Fig. 2-9: Connection cable Pin configuration
Note:
To increase the operational life of this cable, it should not be subject to more than
a minimum amount of movement during operating cycles.
If this plug needs to be removed during fitting, please ensure that this pin
configuration is followed on reassembly.
gr
ye
gn
br
wh
pi
3
1
2
6
5
4