User manual

284
Hand-held Analyzer Types 2250 and 2270 User Manual
D.4 Storage Settings
D.5 Headphone Settings
Table D.4 Storage Settings
Parameter Values Comment
Auto-naming of Projects Yes
No
Select Yes for automatically naming
projects from the start date of the
project as Year, Month, Date in the
format YYMMDD, (for example,
051112 as 2005, November 12th)
Select No for using the name defined
in the Project Name Prefix parameter
Project Name Prefix Text string Prefix for automatically generated
project name. Maximum 8 characters
Table D.5 Headphone Settings
Parameter Values Comment
Listen to signal No
Input X-weighted
Input C-weighted
Input Z-weighted
In addition to the commentary
annotations
, you can listen to the
input signal for monitoring purposes.
Select one of the A-weighted,
B-weighted, C-weighted or
Z-weighted signals.
Note 1: The frequency weighting of
the signal you listen to can be
selected independently of the
frequency weighting of the
measurement, the signal at the
output socket and the signal used
for signal recording
Note 2: X = frequency weighting A
or B. ‘A’ requires that the Broadband
(excl. Peak) parameter is set to AC
or AZ. ‘B’ requires that the
Broadband (excl. Peak) parameter
is set to BC or BZ