Instruction manual

6
Transmitter Cond Ind 7100e/2(X)
Safety precautions for installation
The stipulations of EN 60079-10 / EN 60079-14 must be
observed during commissioning.
The Transmitter Cond Ind 7100e/2H is approved for
measurements in FM Class 1 Div 2.
The Transmitter Cond Ind 7100e/2XH is approved for
operation in the following locations: ATEX, FM Zone 1 with
measurement in Zone 0, and FM Class I Div 1.
Connection to supply units
Transmitter Cond Ind 7100e/2H: Before connecting this
device to a supply unit, make sure that its output voltage
cannot exceed 30 V DC. Do not use alternating current or
mains power supply!
Transmitter Cond Ind 7100e/2XH: This device may only be
connected to an explosion-proof power supply unit (for input
ratings refer to annex of EC-Type-Examination Certificate).
Before commissioning it must be made sure that the
connections to other equipment such as power supply unit
and cables are intrinsically safe.
Terminals:
suitable for single wires / flexible leads up to 2.5 mm
2
(AWG 14)
Note for cleaning in a hazardous location
In hazardous locations the transmitter may only be cleaned with
a damp cloth to prevent electrostatic discharge.
Safety information