Table of Contents I. Safety & Regulatory Notice ............................................................ 2-9 A. Electronic Magnetic Compatibility .................................................. 2-6 B. Safety ............................................................................................... 7-8 C. Intended use.................................................................................... 9-9 II. Product Basic ......................................................................
I. Safety & Regulatory Notice A. Electronic Magnetic Compatibility Guidance and manufacturer’s declaration – electromagnetic emissions The e-medic Silence TP 2 & TP 1 are intended for use in the electromagnetic environment specified below. The user should assure the products are used in such an environment. Emissions test Compliance Electromagnetic environment – guidance RF emissions CISPR 11 The e-medic Silence TP 2 & TP 1use RF energy only for its internal function.
The e-medic Silence TP 2 & TP 1are intended for use in an electromagnetic environment in which radiated RF disturbances are controlled. The user of the e-medic Silence TP 2 & TP 1 can help prevent electromagnetic interference by maintaining a minimum distance between portable and mobile RF communications equipment (transmitters) and the e-medic Silence TP 2 & TP 1 as recommended below, according to the maximum output power of the communications equipment.
discharge (ESD) IEC 61000-4-2 Electrical fast transient/burst ±8 kV air ±8 kV air ±2 kV for power supply lines ±2 kV for power supply lines ±1 kV for input/output lines ±1 kV for input/output lines ±1 kV line(s) to line(s) ±1 kV line(s) to line(s) ±2 kV line(s) to earth ±2 kV line(s) to earth <5 % UT (>95 % dip in UT) for 0,5 cycle <5 % UT (>95 % dip in UT) for 0,5 cycle 40 % UT (60 % dip in UT) for 5 cycles 40 % UT (60 % dip in UT) for 5 cycles 70 % UT (30 % dip in UT) for 25 cycles 70 % U
Power frequency (50/60 Hz) magnetic field 3 A/m 3 A/m Power frequency magnetic fields should be at levels characteristic of a typical location in a typical commercial or hospital environment. IEC 61000-4-8 NOTE UT is the a.c. mains voltage prior to application of the test level. Guidance and manufacturer’s declaration – electromagnetic immunity The e-medic Silence TP 2 & TP 1 are intended for use in the electromagnetic environment specified below.
distance in meters (m). Field strengths from fixed RF transmitters, as determined by an a electromagnetic site survey, should be less than the compliance b level in each frequency range. Interference may occur in the vicinity of equipment marked with the following symbol: NOTE 1 At 80 MHz and 800 MHz, the higher frequency range applies. NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects and people.
B. Safety ISO 7000-0434 : CAUTION, consult ACCOMPANYING DOCUMENTS. ISO 7000-1641 : Follow operating instructions or Consult instructions for use. IEC 60417 -5009 : STAND-BY. IEC 60417-5031 : Direct current. EU-wide legislation, as implemented in each Member State, requires that waste electrical and electronic products carrying the mark (left) must be disposed of separately from normal household waste. This includes monitors and electrical accessories, such as signal cables or power cords.
Collateral standard: Safety requirements for Medical electrical systems – EN 60601-1-2 (IEC 60601-1-2) Medical electrical equipment Part 1-2: General requirements for safety Collateral standard: Electromagnetic compatibility; Requirements and tests Accessory equipment connected to the analog and digital interfaces must be in compliance with the respective nationally harmonized IEC standards (i.e.
C. Intended Use The e-medic Silence TP 2 & TP 1 are designed to serve as a medical PC for general purpose usage in the hospital environment and for clinic review purpose. It could be used as a computer device for Surgical, Radiology, PACS (Picture Archiving Communication Systems), LIS (Lab Information Systems) and Electronic Medical Record application. It shall not be used in life-supporting systems. WARMING: Critical diagnostic decision must not be based solely on images displayed by this device.
II. Product Basic A. Specifications Main System CPU: for TP 2 CPU: for TP 1 Chipset Graphics Memory Serial ATA USB COM Audio LAN BIOS . Support Intel Core i Ivy Bridge ultra low power CPU w/ BGA1023 package, 17W TDP (max.) and integrated graphics . Support Intel Core i Ivy Bridge mobile CPU w/ PGA988 package, 35W TDP (max.) and integrated graphics Intel BD82HM65 PCH Intel HD Graphics HD 4000 8GB maximum, x 2 DDR3 SODIMM, 1066/1333/1600MHz, 2GB/4GB/8GB SATA3 x 2, 600MB/s transfer rate x 7 (socket), USB 2.
Life time 35 million activations Storage & Expansion HDD 2.5” SATA HDD drive bay x 1 + anti-vibration kit Expansion slot . PCI-E (x16) x 1 (optional) . Card size: 100 x 155 x 18/0.5 mm or 68 x 220 x 18/0.5 mm . Mini PCIe x 2 (52 pin card-edge type) I/O & Control Keys USB USB 2.0 x 4 COM DB-9 x 2 (RS232, RS232/422/485) LAN RJ-45 x 2, Gigabit Video-out DVI-I x 1 Audio Line-out, Microphone-in (3.5 mm phone-jack) Power input DC-in x 1, 6V~32V Control keys . Front side, projected capacitive type .
Altitudes Pressure Vibration Shock Regulatory Options Options . Operating: up to 3,000 m (9,842 feet) . Shipping: up to 12,192 m (40,000 feet) . Operating: 700 – 1060 hPa . Storage/Transportation: 186 – 1,060 hPa . Operating: 15g/0.53 oz, 11 ms, half sine wave . Non-operating: 50g/1.76 oz, 11 ms, half sine wave . Operating: 5 ~ 17 Hz , Amplitude - 0.117 ~ 500Hz , Acceleration - 1.0G . Non-operating:10~55Hz/0.15g, 55~500Hz/2.0g rd .
B. The Packing / Outline / I/O / Control Buttons Basic packing list -. The medical PC: e-medic Silence TP 2 (fanless) or TP 1 (smart fan) . Power adaptor (Sinpro MPU101-105) . AC power cord . Touch stylus pen . DVD title (with drivers & user’s manual) . Screws (for rear cover VESA mounting) Note: The contents may vary due to different options or configuration. Outline dimension -.
. e-medic Silence TP 2 .
. The control buttons & LED indicators Control buttons & LED indicators 1. HDD indicator: Dark - non-operating/stand-by Amber blinking - accessing 2. Battery indicator: Dark - non-operating Amber - operating Amber blinking - battery low 3. Power indicator: Dark - off Green - on 4. Power on/off button 5. Speaker volume adjustment (-) button 6. Speaker volume adjustment (+) button 7. LCD brightness (-) button 8. LCD brightness (+) button 9.
C. Others . To clean or disinfect the system To clean or disinfect the system, to avoid any damage to the touch panel and the plastic housing, please check below chemical resistance table to ensure you disinfectant does not contain harmful ingredients. We suggest to use Ethanol and do check with the personal who is responsible for disinfecting.
Chemical Plastic housing 40% Sodium Hydroxid Damage Acetone Damage Carbitol Acetate Damage Cellosolve Acetate Damage Ethanol / Alcohol 500g load with 95% Alcohol--> OK Methylene Chloride Damage Hydrochloric Acid - 6% Damage Hexane Damage Isopropanol (Isopropyl Alcohol)/Propanol Damage Methyl Ethyl Ketone Damage Heptane Damage Toluene 500g load with 5% Cresol --> OK Turpentine Damage Vm & P Naptha Damage Bleach (Sodium Hypochlorite,NaOCI) 500g load with 10% Bleach--> OK Benzyl
. The internal battery (optional) For those systems installed with internal battery, below is the specification. Battery cell type Configuration Typical capacity Minimum capacity Nominal voltage Energy Charge voltage Charge time Battery life Lithium-Ion rechargeable battery cell (2000mAh/cell) 2S2P 4000mAh (for two cells) 3800mAh (for two cells) 7.3 V (average) 29.2 Wh (typical) / 27.74 Wh (minimum) 8.4 V (typical) .System on (1A): 200 minutes .
III. BIOS Immediately after powering on the system, during the POST (Power On Self Test) stage, press the key to enter the BIOS setup utility program. Follow the menu & the keys instruction to modify the BIOS configuration & settings.
Advanced -- Launch OpROM Support Launch PXE OpROM Enables or disables Boot Option for Legacy Network Devices. Launch Storage OpROM Enables or disables Boot Option for Legacy Mass Storage Devices with Option ROM. PCI Subsystem Settings PCI ROM Priority In Case of multiple Option ROMs (Legacy and EFI Compatible), specifies what PCI Option ROM to launch. PCI Latency Timer Value to be programmed into PCI Latency Timer Register. VGA Palette Snoop Enables or disables VGA Palette Registers Snooping.
Extended Tag If ENABLED allows Device to use 8-bit Tag field as a requester. No Snoop Enables or Disables PCI Express Device No Snoop option. Maximum Payload Set Maximum Payload of PCI Express Device or allow System BIOS to select the value Maximum Read Request Set Maximum Read Request Size of PCI Express Device or allow System BIOS to select the value. ASPM Support Set the ASPM Level: Force L0 – Force all links to L0 State : AUTO – BIOS auto configure : DISABLE – Disables ASPM.
Core-Multi Processing Enable or Disable Core-Multi Processing mode. Execute Disable Bit XD can prevent certain classes of malicious buffer overflow attacks when combined with a supporting OS (Windows Server 2003 SP1, Windows XP SP2, SuSE Linux 9.2, RedHat Enterprise 3 Update 3.) Limit CPUID Maximum Disabled for Windows XP. IDE Configuration ATA or IDE Configuration Select ATA or IDE configuration. Configure SATA AS Select a configuration for SATA controller.
TV1 Standard TV2 Standard Active LFP Select the Active LFP Configuration. No LVDS:VBIOS does not enable LVDS. INT-LVDS:VBIOS enables LVDS driver by Integrated encoder. SDV0 LVDS:VBIOS enables LVDS driver by SDV0. USB Configuration Legacy USB Support Allows USB devices to be used in MS-DOS. EHCI Hand-off This is a workaround for 0Ses without EHCI hand-off support. The EHCI ownership change should be claimed by EHCI driver. USB transfer time-out The time-out value for Control, Bulk, and Interrupt transfers.
Serial Port 3 Configuration Set Parameters of Serial Port 3 (COM E). Serial Port 4 Configuration Set Parameters of Serial Port 4 (COM F). Serial Port Console Redirection Serial Port Console Redirection.
Chipset -- Host Bridge/South Bridge This screen provides information on Host Bridge/South Bridge parameters.
Boot -- Setup Prompt Timeout Number of seconds to wait for setup activation key. 65535(0xFFFF) means indefinite waiting. Bootup Numlock State Selects the keyboard NumLock state. Quiet Boot Allows you to determine whether to display the AMI Logo at system startup. Disabled displays normal POST message. Fast Boot Enables or disables the quick boot function to speed up the system boot-up process to shorten the waiting time for entering the operating system and to deliver greater efficiency for daily use.
Always Do not allow disabling GA20. Option ROM Messages Sets display made for option ROM. Interrupt 19 Capture Enables or disables Option ROMs to Trap Int 19. Boot Option Priorities Specifies the sequence of loading the operating system from the installed hard drives.
Security – Enables or disables the security chip. It is recommended that you use this function with the Administrator/User password.
Save & Exit – Save Changes and Exit Exit system setup after saving the changes. Discard Changes and Exit Exit system setup without saving any changes. Save Changes and Reset Reset the system after saving the changes. Discard Changes and Reset Reset system setup without saving the changes. Save Changes Save the changes done so far to any of setup options. Discard Changes Discard the changes done so far to any of setup options. Restore Defaults Restore/load default values for all the setup options.
Baaske Medical GmbH & Co. KG Bacmeisterstr. 3 32312 Lübbecke Tel: +49 5741 236027 – 0 Fax: +49 5741 236027 – 99 vertrieb@baaske.net www.baaske-medical.