Specifications

Legend
1 – on/o buon
2 – numerical display
3 – two piezo speakers (acousc alarm)
4 – upper plasc cover
5 – baery compartment (for two alkaline AA baeries)
6 – green indicator light (sensor pad status), BC-220i and BC-230i have double indicator lights, i.e. green and red combined
in one
7 – red indicator light (alarm status), BC-220i and BC-230i have double indicator lights, i.e. green and red combined in one
8 – control unit
9 – le socket (to connect sensor pads)
10 – right socket (to connect sensor pads)
11 – sensor pad wrapped in plasc
12 – T-connecon (for more pads or for twins)
Pung Baby Control Digital breathing monitor to operaon
(BC-200, BC-210, BC-230)
1. Detach the upper plasc cover (4) from the control unit (8). Parts are not xed with any screw, it is only necessary to put
them apart carefully.
2. Insert two alkaline AA baeries (do not use rechargeable baeries) into the baery compartment (5) inside the control
unit (8). Make sure the baeries are in the marked +/– orientaon.
3. Subsequently, the condion of baeries and control unit (8) are automacally tested. You will hear an acousc signal from
the speaker (3) twice and the green indicator light (6) ashes once and the red indicator light (7) ashes twice.
4. Aach the upper plasc cover (4) to the control unit (8).
5. Connect the sensor pad cable (11) to the socket (9 or 10) located on the side of the control unit (8). The sensor pad (11) can
be connected to any of the sockets (9 or 10).
6. Place the sensor pad (11) evenly at the bed’s grang approximately on the place where the baby’s chest is to be situated
and place the maress over it. Never bend, fold or roll sensor pads!
7. Aach the control unit (8) to the baby’s bed using Velcro clamps. Baby Control Digital cable must be secured so that the
baby cannot pull it. The best way is to turnit around one of the bed bars.
8. Baby Control Digital breathing monitor is ready for operaon.
INFORMATION: When one sensor pad is used (BC-200 type), we recommend placing it under the maress
lengthwise for safety reasons; when two and more sensor pads are used (BC-210 and BC-230 types), place them
crosswise. If all three sensor pads are needed (BC-230 type), connect them to the control unit using the T-con-
necon (part of delivery). If the baby is already able to leave the place, the use of two or three sensor pads is
advised, even though one sensor pad would t the baby. It may occur that the baby moves o one sensor pad
during sleep and so the other pad can take over the sensing funcon.
Typy monitorů dechu Baby Control Digital BC-200, BC-210, BC-230 používejte pouze pro jedno dítě.
U typů BC-220i a BC-230i je možné v případě přesného nastavení (viz REŽIM DVOJČATA) rozdělení senzorových
podložek mezi dvě dě.
IMPORTANT: In rare cases, especially with low weight babies or babies lying on their sides, it may occur that
the sensor pad senses poorly and the displayed measured value is oen below the pre-set threshold level of
detecon. A temporary soluon may be to readjust the threshold level of detecon (see Adjusng the threshold
level of detecon).
Baby Control Digital 6