User guide

Alarm
If the baby is not breathing for longer than 12 seconds, or the breath rate is too slow, after 4
warning signal an alarm signal will sound and the red indicator will flash. On the display(2)
the „H”-„E”-„L”-„P” letters will be displayed. (If movement is detected during the 4 warning
signal the alarm will not sound.
Check the baby and provide first aid if necessary (make sure that there are no reasons for
choking, start resuscitation, call for emergency help).
In some cases of Apnea the sound of the unit itself can wake the baby up and he/she will start
to breath again on his/her own.
Alarm can be stopped by the switching the unit off.
There may be rare cases of false alarms, especially when the child is crawling and moves off
the sensor pad.
E vent Memory
The control unit is capable to store the event of the last 10 minutes before the alarm.
In case of an alarm it is vital to know and understand the previous events for an examination
process.
Review the event memory:
1. Turn off the unit and wait a few seconds
2. Then push the red button(1) and hold it for 10 seconds until you see the „P” and „L”
letters on the display
3. Release the button
4. On the display you can select which minute you would like to see ,„1 P”, „2 P” ... „9 P”
(changes slowly in the display).
5. When you see the required time press the button
6. The measured values will begin to show up
Press the button again and the unit will return to the normal mode.
Battery Replacement
If the batteries are low, the red indicator will flash and the digital display shows error "E",
there will be beep sound after it is tuned on.
For safety reason the unit can not be switched off until you remove/replace the batteries.
In this case replace the batteries. Use only 2 new size AA alkaline batteries (orientation is
clearly marked in the battery compartment). Switch the BABY CONTROL on after battery
replacement and you should hear a beep sound. If there is no beep, check the batteries.
Important: Do not use rechargeable accumulators!