User Guide

Overview Fyrite
INSIGHT
Plus Manual
4 0024-9487 Rev 1
Feature North American (NA) versus Siegert (S) Configurations
Countries
Typical North American
(NA) Users
Typical
Siegert (S) Users
Asia Belgium
Australia Denmark
Latin America France
North America Germany
South America Italy
Netherlands
Poland
Spain
United Kingdom
Heating Values For combustion calculations, Siegert uses the fuel’s lower
heating value; NA uses the higher value (see page 61).
Fuels Different fuel sets and composition (see page 24)
Different RUN
Parameters
EFF (NA) vs. Stack loss and ETA (S)
Excess Air (NA) vs. Lambda (S)
(Lambda is similar to excess air) (see page 61)
Extra Siegert
Parameters
CO/CO
2
ratio, boiler temperature, smoke number, and oil
derivative are displayed for Siegert only (see page 61).
CO
2
Max In the Siegert configuration, the user can set a CO
2
Max number
for the fuel (see page 24).
Print Average
Feature
There is a print average feature for Siegert (see page 31).
Time and Date
Format
NA: MM/DD/YY w/ 12-hour time format with AM/PM or
DD/MM/YY w/ 24-hour time format (see page 47)
Siegert: DD/MM/YY w/ 24-hr time format only
Languages 3 for the North American (NA) configuration and 8 for the
Siegert (S) configuration (See language list on page 44)
NOTE: The Combustion Equations setting is used to configure the
instrument to use either North American combustion equations or
Siegert combustion equations (see page 50). Changing this setting
resets memory and the values of other settings. Refer to page 50
for a list of affected parameters.