User Guide

Fyrite
INSIGHT
Plus Manual Configuration
0024-9487 Rev 1 29
Ambient CO Function
NOTE: Press the F2 key to cancel a test in
progress.
After the test is complete, the Ambient CO Summary
screen is displayed. This is a scrollable window that
shows the 16 CO “snapshot” readings, as well as the
maximum CO reading that was sampled during the
entire test.
NOTE: The Max CO Reading is the highest
sampled CO reading even if the reading
was taken in between one of the sample
“snapshot” readings.
The test results can be printed by pressing F1 and
saved to memory (with a time and date stamp) by
pressing F3. Press F2 to return to the menu.
NOTE: If the ambient CO results are saved
to memory, they are not included as part
of the Print Average feature.
NOTE: Any over-range CO values (e.g., CO
= 4000 ppm) are displayed as “xxx”.