User Guide

Fyrite
INSIGHT
Plus Manual Configuration
0024-9487 Rev 1 45
Setup Menu Function
The analyzer can be set to indicate a calibration
reminder during warmup. Calibration reminders can
be set to occur never, 6, 8, 10, 12, or 15 months after
the last calibration. When the preset period is
exceeded the instrument will display the reminder, and
how long since the sensors were last calibrated. If a
calibration reminder is displayed, the operator can
press the RUN/HOLD key to move to the RUN/HOLD
Screen
for normal operation. Regular calibration
periods of 6 months to 1 year are recommended.
NOTE: The default CAL Reminder Period
is set to NEVER.
Set the calibration reminder period as follows:
1. Use the UP (
) and DOWN (
) arrow buttons to
select the desired time period.
2. Press ENTER to save the selection or ESC to revert
to the previous setting.
NOTE: The date and time settings must
be correct to get accurate calibration
reminders.