User Guide

Fyrite
INSIGHT
Plus Manual Calibration and Maintenance
0024-9487 Rev 1 79
Step CO Sensor Replacement Procedure
5 Plug new CO sensor into its socket.
6 Install the CO cap by aligning it toward the “open” position (12 o’clock) as
shown in the diagram below, then twisting the cap clockwise approximately
40° to the "closed" position (2 o’clock).
7 Reattach tubing.
8 Calibrate the CO sensor (using the calibration procedure on page 88, or using
the B-SMART
procedure on page 80).
O
2
, LL O
2
, and CO Sensor Replacement