User Guide

Fyrite
INSIGHT
Plus Manual Calibration and Maintenance
0024-9487 Rev 1 81
5.5. Pressure Sensor Calibration
This procedure calibrates the pressure sensor to a known pressure value.
5.5.1. Materials Required
Bellows
Manometer
Range: ± 8 in. of water column (± 20 mB)
Accuracy: ± 0.01 in. of water column (± 0.025 mB)
5.5.2. Procedure
NOTE: The unit-of-measure for pressure is selected from the
Pressure Units parameter in the Setup Menu. In the following
procedure inwc” (inches water column) is selected, but note that
any unit-of-measure can be used for calibration purposes. Below
are unit conversions for reference.
249 Pascals/inwc
2.49 mB/inwc
2.49 hPa/inwc
25.4 mm H
2
O/inwc