Specifications

Instruction 19-9220
CO
2
Monitor 2800
2.0 OPERATION
2.1 Reading Mode
When the instrument is switched on, there will be a 20 second (approxi-
mately) warm up time before the CO
2
level present in the atmosphere
surrounding the user is displayed. The ON/OFF button should be pressed
quickly when switching the unit on and held down for approximately 2
seconds when switching the unit off.
CO2 PPM CO2
MONITOR 300
Whe the uit has bee tured o, a “codece bee” will soud for
0.25 seconds every 30 seconds. This should NOT be interpreted as an
alarm condition. The unit has two preset alarm levels which are normal-
ly set to indicate the presence of Carbon Dioxide in concentrations above
5,000 ppm or 10,000 ppm (factory preset levels). The levels can be factory
set at customer secied levels, or ca be reset usig the BACH-COM
Software (See Section 4.1).
When the CO
2
level rises above the low alarm level:
The dislay will show the cocetratio ad ash oce er secod.
The Alarm LEDs will ash oce er secod.
The beeper will sound intermittently once per second.
When the CO
2
level rises above the high alarm level:
The dislay will show the cocetratio ad ash oce er secod.
The Alarm LEDs will be lit continuously.
The beeper will sound continuously.
3