Specifications

CO
2
Analyzer 2825/2830/2835
Instruction 19-9320 7
2.7 Current / Peak Gas Reading Mode
The analyzer can display either the current readings of all measure-
ments, or peak gas readings. While displaying a gas reading, pressing
and holding the f
3
button for 3 seconds switches the analyzer to its peak
reading mode. Note that a peak reading is the highest reading taken
since the analyzer was switched ON.
When using the 2825 / 2830 in its peak-reading mode, pressing the f
3
button toggles the display between ‘peak CO2’ and ‘peak O2’.
To return the analyzer to its current-reading mode, again press and hold
the f
3
button for 3 seconds.
To reset the peak gas readings back to the current readings, switch the
analyzer OFF and then back ON.
2.8 Storing Readings
There are two reading-storage modes available:
• Snapshot
• Continuous Data Logging
Pressing the f
1
button once stores (takes a snapshot of) the reading
shown on the display.
Pressing and holding down the f
1
button for at least 3 seconds starts the
continuous data logging of the readings at a preset interval (factory set
at 30 seconds).
The analyzer can store from 100 to 200 readings based on the analyzer’s
conguration. Logging will stop once the analyzer’s storage capacity has
been reached.
The stored data can be downloaded to a personal computer using the an-
alyzer’s IrDA link and the optional BACH-COM software. This software
can also be used to alter the data-logging interval; produce tables and
plots of time-based CO
2
readings; and clear all logged data from memory.
CO2 ppm peakCO2
340 1,530
ppm CO2
430
Clock icon is displayed once for
a snashot, and ashes during
continuous logging.