Specifications
CO
2
Analyzer 2825/2830/2835
Instruction 0019-9320 1
1.0 INTRODUCTION
1.1 General
Bacharach’s line of CO
2
analyzers are easy to use, but it is essential that
this instruction manual be read, understood, and followed by all operators
and maintenance personnel prior to using or servicing the analyzer.
CO
2
analyzer models 2825, 2830 and 2835 are all compact and light-
weight instruments that principally measure and display CO
2
concentra-
tions, with each of the various models offering the ability to also measure
barometric pressure, O
2
, temperature, and relative humidity.
• 2825 – CO
2
(0–60%), Pressure, and O
2
• 2830 – CO
2
(0–60%), Pressure, O
2
, and Temperature
• 2835 – CO
2
(0–10,000 ppm), Pressure, Temperature,
and Relative Humidity
Each reading is displayed separately on a backlit LCD, with an operator
being able to quickly scroll through all the readings by pressing the f
3
button. A peak-gas-reading mode enables an operator to view the highest
gas reading that was taken since the analyzer was switched ON.
Each analyzer can store all measured readings in memory, which can
later be downloaded to a personal computer via its integral IrDA commu-
nications link and the optional BACH-COM software.
1.2 Features
• Normal and peak reading modes
• Up to 10 hours of operation on one charge
• Convenient zero function for all gas sensors
• Internal pump for remote gas sampling
• Long-life sensors (CO
2
- 10 years; O
2
- 2 years)
• Infrared CO
2
sensor and electrochemical O
2
sensor
• Battery capacity display
• Manual (snapshot) and continuous data logging of readings
with time and date stamp
• Memory capacity for storing 200 sets of data
• IrDA link for downloading stored data to a personal computer
• Charge-and-run capability for long-term monitoring
• Weight: 12 oz