Specifications
CO2 Analyzer 2810/2815/2820
Instruction 19-9222
10
CO2 Analyzer 2810/2815/2820
Instruction 19-9222
11
  2.  Press and hold down the f
2
 button until the ‘User Cal’ screen is dis-
played. Note that battery status is first displayed for approximately 
2 seconds.
 User Cal
  Wait until the analyzer automatically switches to its ‘Air/Gas’ screen 
before proceeding.
 f1-> Air
 f3-> Gas
  3.  Press the f
1
 button to zero the analyzer using fresh ambient air. If the 
air calibration procedure was successful, the display will show
 AirCal
 OK
  If the procedure was unsuccessful, the message 
 AirCal
 Failed
will be displayed. If this happens, retry the Air Calibration procedure, 
ensuring that the analyzer is only exposed to fresh air. If the pro-
cedure is still unsuccessful, then the analyzer must be returned to a 
Bacharach Service Center for evaluation.
  4.  This completes the Air Calibration procedure. If desired the analyzer 
can now be spanned to a known concentration of CO
2
 gas per Section 
3.7.1.2.