Specifications
CO2 Analyzer 2810/2815/2820
Instruction 19-9222
16
CO2 Analyzer 2810/2815/2820
Instruction 19-9222
17
  Wait until the analyzer automatically switches to its ‘N2/Gas’ screen 
before proceeding.
 f1-> N2
 f3-> Gas
  3.  Press the f
3
 button to display the Gas Calibration screen.
 f1->05.0
 00.0
  The top line shows the preset calibration gas value that must be ap-
plied to the analyzer’s inlet. The bottom line shows the current CO
2
reading. 
Note: The analyzer is most accurate around the calibration 
gas value. If the analyzer, however, is routinely used to moni-
tor CO
2
 levels that are much higher or lower than the preset 
value, then it is recommended that the analyzer be calibrated 
on a concentration of gas that is close to the gas level being 
monitored. This operation requires that the preset calibration 
gas value be reset higher or lower using the optional BACH-
COM software, version 2.22 or higher.
  4.  Remove the 100% N
2
 cylinder (if used); then attach a CO
2
 calibration 
gas cylinder to the inlet of the analyzer as shown in Figure 3. The ap-
plied CO
2
 concentration must equal the gas value displayed in the top 
line of the Gas Calibration screen.
Important: Do not attempt to calibrate the analyzer if the 
pressure gauge closest to the tank reads below 5 psi.
  5.  Adjust the regulator attached to the gas cylinder for an outlet pressure 
of 2 psi (equivalent to a flow rate of approximately 120 cc/minute). Ob-
serve that the gas level displayed on the second line of the Gas Calibra-
tion screen should slowly rise to the value displayed in the top line.