Specifications

CO2 Analyzer 2810/2815/2820
Instruction 19-9222
2
CO2 Analyzer 2810/2815/2820
Instruction 19-9222
3
2.3 Pump Operation (2815/2820)
With the analyzer already switched on, momentarily pressing the but-
ton switches on the pump. Pressing the button again turns the pump
off. The supplied sampling probe and hose assembly can be connected
to the gas inlet connector on the top of the analyzer for drawing in gas
samples from hard-to-reach locations.
Particulate filter P/N 54-0548 is placed on top of the gas inlet to prevent
dust or dirt from being drawn into the instrument. Replace the par-
ticulate filter about once a month, or when the filter is contaminated as
evidenced by a slower response time.
Note that the combined sampling hose and probe length should not
exceed 6 feet (1.8 m). Also note that the pump is intended for use only at
normal atmospheric pressure, and is not designed to draw in gas samples
against a vacuum or an obstruction such as a kink in the sampling hose.
If an obstruction or negative-pressure gradient is present, then gas will
not be drawn into the analyzer. Please consult the factory for applica-
tions where longer sampling lengths are required, or where it is neces-
sary to draw against a vacuum.
2.4 Pump Contamination (2815/2820)
Over time the pump can become contaminated, leading to a slower
response time and lower readings. To check for pump contamination,
turn on the pump and hold your index finger over the inlet fitting. The
pump should normally stall. If the pump, however, continues to run,
then the pump is contaminated and the analyzer needs to be returned to
Bacharach for repair.
2.5 High Humidity Gas Sampling (2820)
When using the 2820 to draw in gas samples from areas with high levels
of humidity (e.g., incubators with water jackets), install both particulate
filter P/N 54-0548 and desiccant filter P/N 07-1645 on the analyzer by
first connecting the particulate filter to the analyzer’s inlet. And then
connecting the desiccant filter to the particulate filter using a 2–3 inch
piece of rubber tubing (cut from the tubing supplied with the analyzer).
Use the remaining piece of tubing to connect to the incubator.
After completing a test, continue to run the pump while sampling fresh
air until the reading is below 0.1. This will purge out any residual CO
2
before proceeding to the next test.