Specifications

Instruction 0024-9460
7
Fyrite
®
INSIGHT
General Characteristics:
Power Requirements Four disposable ‘AA’ Alkaline batteries (op-
tional AC adapter is available)
Operating Time Minimum of 20 hours continuous operation
(pump running and backlight ON)
Warm Up Time 60 seconds
Memory Up to 100 complete sets of
combustion, temperature, & pressure
tests can be saved in memory.
Display 128 x 64, LCD Graphic Display
Front Panel Controls Eleven front panel push buttons
(Refer to Section 3.4)
Interfaces IrDA printer & USB connectivity
Accuracy:
Oxygen ±0.3% O
2
with a typical flue gas
concentration of CO
2
Carbon Monoxide ±5% of reading or ±10 ppm, whichever is
greater*, in the range of 0 to 1000 ppm.
Flue Gas Temperature ±4 °F between 32 & 255 °F
(±2 °C between 0 & 124 °C)
±6 °F between 256 & 480 °F
(±3 °C between 125 & 249 °C)
±8 °F between 481 & 752 °F
(±4 °C between 250 & 400 °C)
Ambient/Primary
Temperature
±2 °F between 32 & 212 °F
(±1 °C between 0 & 100 °C)
Draft ±2% of reading or ±0.02 InWC, whichever is
greater in the range of -10 to +10 InWC
± 3% in the range of -10 to -40 InWC
± 3% in the range of 10 to 40 InWC
* Tighter CO accuracy in the lower ranges, up to ±2 ppm, may be
attained if a lower range calibration gas (e.g. 100 ppm CO) is used.