Specifications
Instruction 0024-9460 16
Fyrite
®
INSIGHT
6. Repeat Steps 3-5 to edit the remaining lines.
7.Oncetheinputiscomplete,usethe▲▼buttonstoselectEDIT
COMPLETE.
8. Press ENTER to accept the entered information or ESC to revert to
the previous information and display the SETUP MENU.
Helpful Hint: Use the Fyrite
®
user software to upload this informa-
tion from your computer.
To Clear a Username proceed as follows:
1. Enter the SETUP MENU per Section 3.5.1. If necessary, press ESC
until MENU appears above F2.
2.Usethe▲▼buttonstoselectUSERNAME.PressENTER to dis-
play the most recently saved information.
3. Clear the information by selecting the CLEAR option (F3).
3.5.9 Run/Hold Format
Fyrite INSIGHT test data is located in the Run/Hold screen. By press-
ing the RUN/HOLD key, you should hear the pump running and see the
word RUN at the upper-left hand corner of the display. The instrument
is continuously measuring and calculating the data that is shown in the
Run/Hold screen. Press the RUN/HOLD key again. The pump should stop
running and the word HOLD should be shown at the upper-left hand cor-
ner of the display. The instrument will now show the last measured and
calculated data taken before the instrument was placed in HOLD. Use the
▲▼buttonstoscrollthroughthecompletelistofmeasuredandcalculated
values when the instrument is running or in the HOLD mode.
The order in which data appears in the RUN/HOLD screen is as follows:
O
2
= Oxygen
CO = Carbon Monoxide
Eff = CombustionEfciency
CO
2
= Carbon Dioxide
T-STK = Stack Temperature
T-AIR = Ambient/Primary Air Temperature
EA = Excess Air
CO(n) = Carbon monoxide cont ent referenced to an Ox ygen
percentage
NOTE: (n) is the current O
2
Reference selected.