Specifications
Instruction 0024-9460 18
Fyrite
®
INSIGHT
3.5.10 Language Selection
Information on the display screen can be shown in English, French, or
Spanish.
Select the desired language as follows:
1. Enter the SETUP MENU per Section 3.5.1. If necessary, press ESC
until MENU appears above F2.
2.Usethe▲▼buttonstoselectLANGUAGE
SELECTION. Press ENTER to display the
Language Selection Menu.
3.Usethe▲▼buttonstoselectthedesired
language. Press ENTER to activate the se-
lection and return to the SETUP MENU.
3.5.11 Button Sound
The audible sound used to signal when a button is pressed can be turned
OFF and ON as follows:
1. Enter the SETUP MENU per Section 3.5.1. If necessary, press ESC
until MENU appears above F2.
2.Usethe▲▼buttonstoselectBUTTON
SOUND. Press ENTER to display the Button
Sound Menu.
3.Usethe▲▼buttonstoselectYESorNO
button sound.
4. Press ENTER to save the selection and re-
turn to the SETUP MENU or ESC to revert
to the previous setting.
3.5.12 CAL Reminder Period
The analyzer can be set to indicate a calibration reminder during the 60
second warm-up. Calibration reminders can be preset to occur never, 6,
8, 10, 12, or 15 months after the last calibration. When the preset period
is exceeded the instrument will display the reminder, and how long since
the sensors were last calibrated. If a calibration reminder is displayed the
operator can press the RUN/HOLD key to move to the Run/Hold Screen for
normal operation. Regular calibration periods of 6 months to 1 year are
recommended.