Specifications
Instruction 0024-9460
19
Fyrite
®
INSIGHT
NOTE: ThedefaultCALReminderperiodissettoNEVER.
Set the calibration reminder period as follows:
1. Enter the SETUP MENU per Section 3.5.1. If necessary, press ESC
until MENU appears above F2.
2.Usethe▲▼buttonstoselectCALRE-
MINDER PERIOD. Press ENTER to dis-
play the Cal Reminder Period Menu.
3.Usethe▲▼buttonstoselectthedesired
time period.
4. Press ENTER to save selection and return to
the SETUP MENU or ESC to revert to the
previous setting.
NOTE: The date and time settings must be
correct to get accurate cal reminders.
3.5.13 Inactivity Timeout
The analyzer can be placed in an inactivity timeout period of None, 30 or
60 minutes. With this time period set the analyzer will automatically shut
off if there have been no button pushes made in the set period of time. The
analyzer will not turn off if the CO value is greater than 50 ppm or the O
2
value is less than 16%.
NOTE: The default set time is NONE which prevents the analyzer
from automatically turning off.
Set the Inactivity Timeout Period as follows:
1. Enter the SETUP MENU per Section 3.5.1. If necessary, press ESC
until MENU appears above F2.
2.Usethe▲▼buttonstoselectINACTIVITY
TIMEOUT. Press ENTER to display the
Inactivity Timeout Menu.
3.Usethe▲▼buttonstoselectthedesired
timeout period.
4. Press ENTER to save the selection and re-
turn to the SETUP MENU or ESC to revert
to the previous setting.