Specifications

Instruction 0024-9460 36
Fyrite
®
INSIGHT
• Oxygen sensor spanned to 20.9%
• CO sensor zeroed
• Pressure sensor zeroed.
Any errors detected during warm-up will be listed on the dis-
play immediately following warm-up. Correct any errors before
proceeding. Refer to Section 6.8 for a list of error messages and
their meaning.
3. Display the MAIN MENU by pressing the MENU (F2) button. If
necessary, press ESC until MENU appears above F2.
4.Usethe▲▼buttonstoselectCALIBRA-
TION. Press ENTER to display the Calibra-
tion Password screen.
5. Before calibration can begin a 4-place nu-
meric password must be entered. Use the
▼and◄►buttonstoenterthepassword.
(Note that the default password is 1111).
6. Press the ENTER button to accept the pass-
word. If the correct password was entered
the CALIBRATION Menu will be displayed.
7.Usethe▲▼buttonstoselectthedesiredsensorchanneltobe
calibrated, and then perform the calibration procedure for that
sensor as described in the following sections.
5.3 B-Smart Sensor Replacement & Calibration
To replace a B-Smart sensor and calibrate do the
following:
1. Enter the CALIBRATION MENU per Sec-
tion 5.2.
2.Usethe▲▼buttonstoselectB-SMART.
Press ENTER to display the B-Smart code
screen.
3.Usethe▲▼buttonstoenterthe10digit
alphanumeric code supplied with the B-
Smartsensor.Usethethe◄►buttonstomovethecursoracross
the screen. Press ENTER.