Specifications

Instruction 0024-9460
37
Fyrite
®
INSIGHT
NOTE: If the correct code was entered, the analyzer accepts it and
returns to the CALIBRATION MENU. If an incorrect code
was entered, the screen will display "Invalid Code". Check
to make sure the correct code has been entered. If problem
persists, contact your nearest Bacharach Service Provider.
Bacharach also offers a convenient Exchange Program that allows the cus-
tomer to regularly send in old sensors and, as scheduled, receive new re-
placements that are already calibrated and that include a code that can be
entered into the analyzer for a quick convenient setup. Contact Bacharach
customer service for more details about this program.
5.4 Pressure Sensor Calibration
This procedure calibrates the pressure sensor to a known pressure value.
Materials Required:
• Bellows
• Manometer
- Range: ± 8 in. of water column (± 20 mb)
- Accuracy: ± 0.01 in. of water column (± 0.025 mb)
Procedure:
NOTE: Theunit-of-measureforpressureisselectedperSection3.5.3.In
the following procedure inwc is selected, but note that any unit-
of-measure can be used for calibration purposes.
Fig.5-1PressureSensorCalibrationEquipment