Specifications

Instruction 0024-9460 42
Fyrite
®
INSIGHT
TA-Span Procedure:
6. Set thermocouple simulator to 212 °F (100 °C),
and then use the and buttons to enter
an Applied value that exactly equals the set-
ting of the simulator.
Alternatively: Submerge probe tip into a container of boiling water
with a thermometer, wait several minutes, and then use the and
buttons to enter an Applied value that exactly equals the ther-
mometer reading.
Thecalibrationrangeisfrom194to230°F(90to110°C).Anat-
tempt to calibrate outside this range will cause the message “Bad
CalibrationWrongCALEntry”toappearinthefollowingstep.
7. Wait until the Measured reading stabilizes, and then press ENTER
to calibrate the TA-Span Measured value to that of the Applied
value; after which the message “Good Calibration” should briefly
appear followed by the CALIBRATION LIST screen being re-dis-
played.
5.7 CO Sensor Calibration
Material required:
• Calibration kit, P/N 0024-7059
• Gas cylinder: 500 ppm CO in air, P/N 0024-0492
To improve the accuracy of the CO reading, we suggest that if the analyzer
will be primarily used for flue gas testing, then calibrate using 500 ppm
CO. If the analyzer, however, will be primarily used for ambient testing,
then calibrate using 100 ppm CO.
Procedure:
1. If not already done, turn ON the analyzer
and display the CALIBRATION LIST
screen per Section 5.2.
2. Use the buttons to highlight CO, and
then press ENTER to display the CALI-
BRATE CO screen.