Specifications
Instruction 0024-9460
53
Fyrite
®
INSIGHT
Bad Sensor - O
2
sensor is too low and can not be calibrated in the instru-
ment and needs replaced.
Low Sensor- O
2
or CO sensor outputs were low but still usable. Sensor(s)
may need to be replaced in the near future. Message will indicate which
sensor(s) were in warning.
Warmup Sensor Error -
•COsensorwasnotzeroedatwarmupbecauseofhighoutput.Run
instrument on fresh air then restart instrument to re-zero sensor. If
message persists, CO sensor may need to be replaced.
•StackorAirtemperaturechannelismeasuringtemperatureoutside
therangeof-4to212˚Fatstartup.MakesurethattheStackand
Air thermocouples are sampling ambient room air within the tem-
perature range at startup.
•Pressuresensorismeasuringpressureoutsidetherangeof±3inwc
at startup. Ensure that the analyzer is sampling atmospheric pres-
sure and restart the instrument.
•Theanalyzerwasturnedonwiththeprobesamplingfluegas.Move
the probe to fresh air and restart the instrument.
•Messagewillindicatewhichchannelisinerror.
Low Battery - Battery voltage is low. Replace the batteries.
XXXX-Occursinthenumbereldsofsensorsthathaveachievedover-
range condition.
****-Occursinthenumbereldsofsensorsandthecalculatedvalues
that depend on the sensors that were in error coming out or warmup.
- - - - Occurs in the number fields of calculated values when Oxygen is
above 16.0%.
6.9 Diagnostics Screen
The DIAGNOSTICS menu provides information regarding the operation of
the analyzer. Information includes the following:
Time Meters - Displays the run time of the analyzer and sample pump in
hours of operation.
Main Diagnostics - Lists the current status of the Stack and Air Thermo-
couple channels, Reference Temperature channel, Pressure channel, CO
channel, O
2
channel and Battery.
O
2
Sensor Life - Displays the approximate remaining life of the Oxygen
sensor.